4-9
Processing Instructions
Operator Manual
920011-649
b. Carefully feed distal end of insertion tube into insertion tube
slide
.
Feed entire insertion tube
into slide.
IMPORTANT: Hold insertion tube near control handle boot
opening to ease insertion and prevent damage caused by
kinking or excessive bending.
NOTE:
For insertion tubes that are hard to insert into tube
slide, wet tube with water to ease insertion.
12.
If using split insertion tube sleeve (NOT labeled with a color
dot),
proceed as follows (see
:
a. Open split insertion tube sleeve as shown in Figure 4-8.
CAUTION –
POSSIBLE EQUIPMENT
DAMAGE: Do not place light
guide/ultrasound connectors
on their cables or damage
may result.
Figure 4-6. Insert Insertion Tube Into Control
Handle Boot Opening
IMPORTANT:
Hold Insertion Tube Near Control Handle Boot
Opening to Prevent Excessive Bending or Kinking.
Insertion Tube
Control
Handle
Light Guide Cable
Control Handle
Boot Opening
Figure 4-7. Place Control Handle in Control
Handle Boot
Universal
Cord Port
Flaps
Light
Guide
Cable
Strain
Relief
Boot
Insertion
Tube
Slide
Endoscope Control
Handle
Figure 4-8. Split Insertion Tube Sleeve
WITHOUT a Color
Identification
Opening Split Insertion Tube Sleeve
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...