985-006001 REV. G Effective Date: 03 June 2008
Stereotaxis, Inc.
Page 9 of 43
3.2. System Power On/Off and Emergency Stop
NOTE
If an X-ray system is not available at this time use the X-ray simulator for testing purposes.
No.
Test Description
Pass/Fail
Initials
1
Verify that the Emergency Stop buttons function
•
While the system is powered on and operational depress the
emergency stop button in the control area and verify that the
red light labeled “E Stop Active” behind the left hand front door
of the Control cabinet illuminates and all motion of the MPs is
inhibited.
•
While the system is powered on and operational depress the
emergency stop button on a tableside controller and verify that
the red light labeled “E Stop Active” behind the left hand front
door of the Control cabinet illuminates and all motion of the
MPs is inhibited.
P F
P F
2
Verify that the System/Positioner power on/off button functions
•
While the system is powered on (the system/Positioner power
on light is lit) actuate the on/off switch by holding the button in
the OFF position (approx. 3 sec) until power is removed.
Verify that the on/off light is no longer lit, the red light labeled
“E Stop Active” behind the right hand front door of the Control
cabinet and the System Control/Remote Stop Unit** are
extinguished, and all motion of the MPs is inhibited.
•
While the system is powered off (the system/Positioner power
on light is not lit) actuate the on/off switch to the on position.
Verify that the on/off light illuminates, the red light labeled “E
Stop Active” behind the right hand front door of the Control
cabinet and the System Control/Remote Stop Unit** are
extinguished, motion of the MPs is restored.
P F
P F
System/Positioner
power on when
illuminated (Green)
System/Positioner
power on/off
Emergency Stop
is activated when
illuminated (Red)
Emergency Stop
(E-Stop). Press to
halt all mechanical
motion.
Figure 1 – System Control / Remote Stop**
(001-005000-1 and 001-006000-1 configuration shown)