17
Installing and Removing a Sensor
SensoGate® WA 130 H Retractable Fitting
Sensors must only be installed or removed by trained personnel authorized by the
operating company.
Make sure that the retractable fitting is in SERVICE position
(see “Function Description” on Page 10)
.
Be sure to follow the assembly instructions step by step.
Preparations:
• Check whether the sensor is damaged (glass broken?).
Never install a damaged sensor.
• Check whether slide washer or O-ring on the sensor are damaged and replace if required.
• Remove watering cap from the sensor tip and rinse sensor with water.
• Internally pressurized sensors might have a silicone seal on the diaphragm (as transport
protection). Remove this seal using the knife shipped with the sensor.