24
Removing the Drive Unit
Step-by-Step Instructions
CAUTION
:
Be sure to follow the steps below in the correct order.
Take appropriate safety precautions against escaping process fluids.
1) Move probe into SERVICE position.
2) Make sure that no process fluid is leaking from the outlet
(I)
.
3) If required, remove sensor as described in section “Installing and Removing a Sensor” on Page
17.
4) Separate outlet
(I)
and rinse connection if required.
5) Turn coupling nut
(B)
counterclockwise (using the ZU 0680 accessory wrench if required –
see figure). Do not cant the unit and do not exert force.
6) Pull off the drive unit upwards
(E)
.
I
B
E