22
Removing a Liquid-Electrolyte Sensor
Removing the sensor
1) The sensor must only be removed in
SERVICE
position
.
2) Remove cable socket with cable.
3) Before removing the sensor, check that there is
no liquid leaking from the outlet (process sealing
might be defective).
4) Loosen small coupling nut
(R)
– do not remove
it.
5) Unscrew large coupling nut
(S)
completely and
pull the detached unit upwards.
6) Remove sensor
(E)
.
R
S
E