Page 14 of 54
Pro
Top Level
(Programming menu entry point)
CnInp
Level 1
Channel Input; for selecting the correct type of input configuration for the sensor
to be used.
Options available; Proc, LoAd, tEMP
Proc
Level 2
(See Sec. 3.1)
LoAd
Level 2
(See Sec. 3.2)
Process input type selection.
Load cell input selection
Options available; 10 V, 20 mA
Options available; 15 mV, 30 mV,
150 mV
10 V
Sub level
15 mV
Sub level
Input range +/ (0 to 10) V
Input range (-15 to 15) mV
20 mA
Sub level
30 mV
Sub level
Input range +/ (0 to 20) mA
Input range (-30 to 30) mV
150 mV
Sub level
Input range (-150 to 150) mV
Temp
Level 2
Temperature input type selection. Options available are PT100 and TC
Pt100
Level 3
See 3.3
tC
Level 3
See 3.4
Pt100 input type configuration,
options available are:
Display unit, Display resolution, User
offset
Thermocouple input type
configuration, options available are:
TC type, Display unit, Display
resolution, User offset
J, K, T
Sub level
TC type selection.
°
C or °F
Sub level
The displayed temperature unit can be selected.
0.1° or 1°
Sub level
The display resolution can be selected.
Offset
Sub level
A value can be entered by the user which will be applied to the measured value
before it is displayed. The value entered is applied across the entire measurement
range. This can be used to tune out system errors.