6. PROGRAMMING LOCK OUT BY
SOFTWARE
The instrument is delivered with access to all the programming
levels. Once the instrument is configured, it is possible to restrict
access to the configuration menus
1.
There are two lockout modes: selective and total. If the
parameters are going to be readjusted frequently, make a
selective lockout. If no adjustment is going to be made, make
a total lockout.
2.
Access to the lockout routine is allowed by entering a
personalised code. We recommend changing the code set at
the factory and to write down your personalised code and
keep it in a safe place.
TOTAL LOCKOUT
Access to the programming routines to read data is allowed even if
all parameters are locked out totLC=1, but
it won’t be possible
to enter or modify data
. In this case, when entering in the
programming mode, the display shows the indication "-dAtA-".
PARTIAL LOCKOUT
When only some parameters are locked out, all configuration data
can be read but
only non-protected parameters can be
modified
. In such case, when entering in the programming mode,
the display shows the indication “
-Pro-".
Menus or submenus that can be locked out are:
•
Setpoint 1 configuration (SEt 1).
•
Setpoint 2 configuration (SEt 2).
•
Setpoint 3 configuration (SEt 3).
•
Setpoint 4 configuration (SEt 4).
•
Input configuration (InPut).