Getting Started
Battery
To insert a battery, open the compartment by sliding the door to the right and remove the door
from the unit. Insert the battery, with the terminals positioned as shown in the diagram on the
floor of the battery compartment.
Connecting the Pick-up
Plug the lead into the socket on the front of the unit and mount the pick-up as follows: The
connector of the pick-up lead is screwed into the end of the pick-up and is then inserted into
the end of the pick-up holder, with the lead coming out through the slot in the holder. It is
advisable to connect the lead to the display-traverse unit first and then the pick-up. To connect
the pick-up to the dis-play-traverse unit: the pick-up has 2 threaded ends with location pins.
Insert the location pin securely into the SR200 body and tighten the threaded collar.
Rotate the pick-up to bring the stylus vertical; this can conveniently be done with reference to
the identity number engraved at the end of the pick-up.
Position the Pickup stylus on the component to be measured, with the stylus parallel to the
component (see figure 8 below). Ensure that the stylus tip is in contact with the surface (this
can be verified by checking that the identity number engraved at the end of the pick-up is
vertical). The skid should also be in contact with the surface.
Figure 8
Rapp Industrial Sales 724 789-7853