5.2
SR200
Specifications subject to change
Cut off
A few trial measurements made on different surfaces will soon demonstrate that on some, the
results obtained are very dependent upon the cut-off selected. This shows that it is important
to choose the cut-off to suit the surface. In general, fine surfaces require short cut-offs and
rough surfaces a longer one. The table on the following page gives some guidance on suitable
cut-offs.
Evaluation Length
A long evaluation length is mainly for use on sheet metal and similar materials, where a longer
surface is required to be representatives of the material being examined.
Table of Cut-off values
If not otherwise indicated on a drawing, the following should be used to determine the cut-off
λ
c (ISO4288)-1996. Recommended Cut-off ISO 4288-1996
Periodic Profiles
Non-periodic Profiles
Cut-offs
Sampling Length/
Evaluation Length
Spacing Distances
Sm (mm)
Rz(µm)
Ra (µm)
λ
c (mm)
λ
c/L (mm)
>0.013 to 0.04
>0.04 to 0.13
>0.13 to 0.4
>0.4 to 1.3
>1.3 to 4
(0.025) to 0.1
>0.1 to 0.5
>0.5 to 10
>10 to 50
>50 to 200
(0.006) to 0.02
>0.02 to 0.1
>0.1 to 2
>2 to 10
>10 to 80
0.08
0.25
0.8
2.5
8
0.08/0.4
0.25/1.25
0.8/4
2.5/12.5
8/40
Operating Error Indications
During a measurement a message may be displayed which indicates that an error condition has
occurred. The messages and the probable causes for their display are as follows:
Display
Reason for message
Motor Error
Motor fault
Pick-up Error
Response at pick-up level error or faulty connection
Over range
Selected range too small
E
Can occur in front of a measured value due to excessive slew rate for
a single data point (large spike). May also occur if the pick-up is lifted
off the surface
Data transmit error
Printer connected at start of transmission but later detected as not
connected (printer off line)
Battery low
When battery voltage <6.4V: key push operations are disabled and
display is off.
When the instrument is in the sleep mode and the MEASURE
key
is pressed, then “Battery low” is displayed for 2 seconds, followed by
the normal display or operation.
Stored surface profile is erased
Renew or recharge battery