 Recommended separation distances between portable and mobile RF communications equipment and the NuTorque Lite

The NuTorque Lite is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the 

NuTorque Lite can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment 

(transmitters) and the NuTorque Lite as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter

W

Separation distance according to frequency of transmitter

150kHz to 80MHz

d=1.2   P

80MHz to 800MHz

d=1.2   P

800MHz to 2.5GHz

d=2.3   P

0.01

0.12

0.12  

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the 

equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter 

manufacturer.
NOTE 1  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Conducted RF

IEC61000-4-6

Radiated RF

IEC61000-4-3

3Vrms

150 kHz to 80MHz

3V/m

80MHz to 2.5 GHz

3Vrms

3V/m

Portable and mobile FRO communications equipment should be used no closer 

to any part of the NuTorque Lite, including cables, than the recommended 

separation distance calculated from the equation applicable to the frequency of 

the transmitter.

Recommended separation distance

d = 1.2  P
d = 1.2  P  80MHz to 800MHz

d = 2.3  P  800MHz to 2.5GHz

Where 

P is the maximum output power rating of the transmitter in watts 

(W)according to the transmitter  manufacturer and 

d is the recommended 

separationdistance in meters (m).

Field strengths from fixed RF transmitters as determined by an electromagnetic 

site survey

a

, should be less than the compliance level in each frequency range

b

Interference may occur in the vicinity of equipment marked with the following 

symbol:

  

NOTE 1  At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobiles radios, amateur radio, AM and FM radio 

broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 

electromagnetic site survey should be considered. If the measured field strength in the location in which the NuTorque Lite is used exceeds the applicable RF 

compliance level above, the NuTorque Lite should be observed to verity normal operation. If abnormal performance is observed, additional measures may be 

necessary, such as reorienting or relocating the NuTorque Lite .
b  Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m.

Guidance and manufacturer's declaration - electromagnetic immunity  
The NuTorque

®

 Lite is intended for use in the electromagnetic environment specified below. The customer or the user of the NuTorque Lite and manual should 

assure that it is used in such an environment.

Immunity test

IEC60601 test level

Compliance level

Electromagnetic environment - guidance

Electrostatic discharge (ESD) 

IEC61000-4-2

±6kV contact

±8kV air

±6kV contact

±8kV air

Floors should be wood, concrete or ceramic tile. If floors 

are covered with synthetic material, the relative humidity 

should be at least 30%.

Electrical fast transient/burst 

IEC61000-4-4

±2kV for power supply lines

±1kV for input/output

±2kV for power supply lines

No input/output line

Mains power quality should be that of a typical 

commercial or hospital environment.

Surge IEC61000-4-5

±1kV line(s) to line(s)

±2kV line(s) to earth

±1kV line to line

±2kV lines to earth

Mains power quality should be that of a typical 

commercial or hospital environment.

Voltage dips, short 

interruptions and voltage 

variations on power supply 

input lines IEC61000-4-11

<5% Ut

(>95% dip in Ut) for 0.5 cycle

40% Ut

(60% dip in Ut) for 5 cycles

70% Ut

(30% dip in Ut) for 25 cycles

<5% Ut

(>95% dip in Ut) for 5 sec 

<5% Ut

(>95% dip in Ut) for 0.5 cycle

40% Ut

(60% dip in Ut) for 5 cycles

70% Ut

(30% dip in Ut) for 25 cycles

<5% Ut

(>95% dip in Ut) for 5 sec

Mains power quality should be that of a typical 

commercial or hospital environment. If the user of the 

NuTorque Lite requires continued operation during power 

mains interruptions, it is recommended that the NuTorque 

Lite be powered from an uninterruptible power supply or 

a battery.

Power frequency (50/60Hz) 

magnetic field IEC61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels 

characteristic of a typical location in a typical commercial 

or hospital environment.

NOTE: Ut is the a.c. mains voltage prior to application of the test level.

Содержание NuTorque Lite

Страница 1: ...plains an instruction that should be observed for safety reasons Read this manual thoroughly before use and operate the device properly These instructions will show you how to operate the product safety and prevent danger to you or others They are classified by the degree and or severity of danger All contents relating to safety should be oberserved Cautions for handling and operation Please Read ...

Страница 2: ...the Speed Control Switch is adjusted within the allowable speed before use When you install Control Box or Motor DO NOT bend or twist the tubing or the cord No unnatural force is needed to do the installation Never use the AC adapter for anything other than the control box of this product Air Requirements dry free from contamination and oil Use a compressor with a dry air system Install an air fil...

Страница 3: ...Model 265290 120V and 265291 230V Rated input AC120V 50 60Hz 41VA AC230V 50 60Hz 41VA Output AC28V 1 3A Fuse AC120V FU100 T1 6A 250V FU101 T1 6A 250V AC230V FU100 T800mA 250V FU101 T800mA 250V Dimensions W100 x D178 x H64 mm Cord Length Unshielded App 5 0m AC Adaptor Handpiece Tubing Connector Handpiece Tubing Connector Power Switch Power Switch Power Switch AC Adapter Connector AC Adapter Connect...

Страница 4: ...e Motor Insert the Motor Cord Connector to the control box motor cord connector and tighten Fig 2 3 Connecting the AC adapter Insert the AC adapter plug into the AC adapter connector before plugging into wall outlet Fig 3 Never use the AC Adapter for anything other than the Control Box of this product 4 Connecting the Power Cord Insert the Power Cord into the Inlet of AC Adapter Fig 4 Plug cord in...

Страница 5: ...the Power Switch When you turn on the Power Switch lamps and speed indicator will be illuminated 2 Before using this system make sure to perform Foot Air Calibration function Refer to 6 1 Foot Air Calibration Power Switch Symbol Mark Function OFF ON 3 Push the Gear Ratio Select Key to indicate the speed of the attachment you will use Fig 7 Handpiece GearRatio Indicator Speed min 1 rpm 16 1 100 2 5...

Страница 6: ...wer limit of the air pressure to enable this function Perform this adjustment function after your purchase Once it is adjusted it will be memorized and there is no need to do this everytime NOTICE The displayed air pressure is only a guideline Measure the correct air pressure with an air gage The unit of displayed air pressure is kgf cm2 For other units refer below ex 0 1kgf cm2 0 01MPa 1 42psi 1 ...

Страница 7: ...table between 4 Press key until it beeps more than 3 seconds to set the brightness 5 Press key and key at the same time for more than 3 seconds once again When it beeps this setting is complete 3 Program Function It is possible to program two 2 presets Speed Gear Ratio Forward Reverse Direction Once set you can choose between presets by pressing the or the key To set the values for Speed Gear Rati...

Страница 8: ...t from motor 5 Disconnect motor from tubing by unscrewing coupler counter clockwise and pulling motor in the axial direction direction of E connector 6 Clean all external surfaces of attachment and motor using a small toothbrush and soft cloth with isopropyl alcohol 7 Dry thoroughly lubricate attachment 8 Sterilize Note To Prevent Damage To the Bur Use an Autoclavable Bur Block Attachments and bur...

Страница 9: ...d mount the new O Rings into the groove CAUTION Effects of deteriorated O Rings Air Water leak Air Water does not turn off Vibration Difficult to connect disconnect the handpiece 2 Replacing Bulb 1 Turn off the power 2 Loosen Motor cover and detach it from the motor Remove the old light bulb using a small screw driver Fig 12 13 3 Align and insert the connector of the new bulb into the socket hole ...

Страница 10: ...setting values are different from what used to be when turned off Turned off the motor while rotating Turn off after the Motor has stopped 9 Error Code Error Code Trouble Cause Check Remedy Overcurrent error In circuit Detected overcurrent in the circuit Contact your dealer Overcurrent error In software Detected overcurrent in the software Contact your dealer Retry Error Motor stopped by overload ...

Страница 11: ...orque Lite uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emmissions CISPR11 class B The NuTorque Lite is suitable for use in all establishments including domestic estabilishments and those directly connected to the public low voltage power supply network that supply network that s...

Страница 12: ...a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobiles radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the locatio...

Страница 13: ...l information call 1 866 DTE INFO or visit our web site www dentalez com StarDental Division of DentalEZ Inc 1816 Colonial Village Lane Lancaster PA 17601 U S A Printed in Japan 2019 DentalEZ Inc Anyone wishing to return any DentalEZ Inc product to a DentalEZ facility MUST obtain the proper RETURN AUTHORIZATION NUMBER from the Customer Service Department The RA Number must appear clearly marked on...

Отзывы: