4.5 Commands
4 – 35
4.5.11
TEC monitor commands
TEC Current Reading
TIRD?
Query the TEC operating current, in A.
TIRD?
always returns the operating current flowing to the TEC mod-
ule, independent of TEC control mode (CT or CC).
Front panel:
(@HHH*
.
TIRD?
Example:
-1.00
TEC Voltage Reading
TVRD?
Query the TEC voltage, in V.
Front panel:
GH
.
TVRD?
Example:
1.00
Raw Thermometer Reading
TRAW?
Query the raw sensor value, in k
Ω
, V, or
μ
A (depending on the sensor
type).
Front panel:
3HG01F
.
TRAW?
Example:
1.01
Celsius Thermometer Reading
TTRD?
Query the sensor temperature, in
◦
C.
Note that
TTRD?
is not legal, and will generate an execution error,
when the sensor model is configured to “none” (resistance mode).
Front panel:
IH
d
.
TTRD?
Example:
2.01
LDC500 Series Laser Diode Controllers