4.5 Commands
4 – 33
Note that the temperature limits are not e
ff
ective if the sensor model
is set to “none” (resistance units).
Front panel:
"#
E G)
d
.
TMAX 65.0
Example:
Lower Resistance Limit
TRMN(?)
{
f
}
Set (query) the lower resistance limit
{
to
f
}
, in k
Ω
.
When the LDC500 is operated with the sensor model set to “none”
(resistance units), then
TRMN
establishes the minimum sensor resis-
tance for normal operation. If the resistance goes below
TRMN
, and
TTMN
is set, then the TEC shall turn o
ff
; if
ATMN
is set, then the LD
shall turn o
ff
.
Note that
TRMN
establishes the lower bound on resistance, regard-
less of whether the sensor has a negative or positive temperature
coe
ffi
cient.
Front panel:
"#
3 H)1F
.
TRMN 9.50
Example:
Upper Resistance Limit
TRMX(?)
{
f
}
Set (query) the upper resistance limit
{
to
f
}
, in k
Ω
.
When the LDC500 is operated with the sensor model set to “none”
(resistance units), then
TRMX
establishes the maximum sensor resis-
tance for normal operation. If the resistance goes above
TRMX
, and
TTMX
is set, then the TEC shall turn o
ff
; if
ATMX
is set, then the LD
shall turn o
ff
.
Note that
TRMX
establishes the upper bound on resistance, regard-
less of whether the sensor has a negative or positive temperature
coe
ffi
cient.
Front panel:
"#
3E )1F
.
TRMX 12.50
Example:
4.5.10
TEC setting commands
LDC500 Series Laser Diode Controllers
Содержание LDC500
Страница 14: ...xii General Information LDC500 Series Laser Diode Controllers...
Страница 50: ...2 22 Getting Started LDC500 Series Laser Diode Controllers...
Страница 128: ...4 64 Remote Operation LDC500 Series Laser Diode Controllers...
Страница 136: ...5 8 Error Messages LDC500 Series Laser Diode Controllers...