Model SR575 Bariatric Stand-On Scale
Operating and Service Manual - S/N 1000+
Part No. MAN575_031105
Page 9 of 18
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
2003
S
Instruments, Inc.
BASIC SYSTEM OPERATION Cont’d
AUTO-HOLD
This mode is for patients unable to remain still for the weighing procedure. It locks, stores, and
displays the patient’s weight as soon as the “
WEIGH
” button is pressed once.
Note
: No weight
will be displayed until the button is pressed.
To enable this mode, before zeroing the system, press and hold the “
HOLD
” button for
approximately five (5) seconds until the display reads “
AUTO-HOLD ENABLED
”.
To return to CONTINUOUS WEIGH mode when finished, press and hold the “
HOLD
” button for
approximately five (5) seconds until the display reads “
CONTINUOUS WEIGH
”.
BATTERY REPLACEMENT
STEP 1
: The display will read “
REPLACE BATTERY
”.
STEP 2
: (Figure 4) Tip the scale on its side and unscrew the two (2) screws from the Battery
Compartment Cover located on the base of the unit, between the wheels
.
STEP 3
: Remove and replace ALL six (6) “D” cell batteries. Refer to Battery Compartment Cover
label for placement.
STEP 4
: Press the “
WEIGH
” button to confirm display is working.
STEP 5
: Tightly close the cover.
STEP 6
: Zero the system.
Figure 4: Battery Compartment