
Model SR575 Bariatric Stand-On Scale
Operating and Service Manual - S/N 1000+
Part No. MAN575_031105
Page 11 of 18
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
2003
S
Instruments, Inc.
LOW
LIMIT
APPLIED
LOAD
HIGH
LIMIT
99.9
100.0
100.1
199.8
200.0
200.2
299.7
300.0
300.3
399.6
400.0
400.4
499.5
500.0
500.5
599.4
600.0
600.6
699.3
700.0
700.7
799.2
800.0
800.8
899.1
900.0
900.9
999.0
1000.0
1001.0
CALIBRATION
TOLERANCE TABLE
CALIBRATION
Note:
Ensure that nothing is in contact with the scale system during this procedure. Remove
hands from the system when noting the displayed calibration results.
STEP 1
: (Figure 5) Simultaneously press and hold the hidden calibration buttons on the label
(“
+
” and “
-
”)
.
The display will read “
HOLD TO CAL
” as the right hand digit counts down from
9 to 0 to enter the CAL mode.
STEP 2
: When in the CAL mode, press the “
ZERO
” button to zero
the display.
STEP 3
: Place a known calibrated weight, traceable to NIST, onto the
weighing surface and compare it to the displayed reading.
Note
: DO
NOT USE barbell weights or calibrate to a mechanical scale
.
STEP 4
: Use the “
+
” and “
-
” buttons to make any necessary
adjustments to the displayed value. The displayed value should
be within 0.1% of the calibrated weight, plus or minus 1 digit of
reading.
STEP 5
: When adjustments are completed: Press the “
HOLD
”
button to SAVE the settings, or press the “
WEIGH
” button to
CANCEL. Both choices will EXIT the CAL mode.
Figure 5: Location of Hidden Calibration Buttons
CAUTION
ESD: The integrated circuits and
semiconductors on the printed circuit
boards may be damaged by electrostatic
discharge (ESD). Be sure to use proper
handling precautions at all times.
IMPORTANT
CALIBRATION CHECK - Qualified service personnel only should perform this
procedure. Load cells have no user serviceable components and should not be tampered
with for any reason. Re-calibration is generally not required, but should be verified
periodically to ensure accuracy. Recommendation for calibration check is at least once
every 12 months, or as individual maintenance policy requires.
HIDDEN CAL BUTTONS
+ _