Model SR575 Bariatric Stand-On Scale
Operating and Service Manual - S/N 1000+
Part No. MAN575_031105
Page 15 of 18
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
2003
S
Instruments, Inc.
ER44991EJ
49.9K 1/8 WATT 1% J RESISTOR
R8
ER54991EJ
499K 1/8 WATT 1% J RESISTOR
R16
ER55491ED
549K 1/8 WATT 1% RESISTOR
R15
ER62004EJ
2M 1/8 WATT 1% J RESISTOR
R5,R14
ER72205QF
22M 1/4 WATT 5% RESISTOR
R9
ER770101
RESISTOR NETWORK 100K
RSIP1,2,3,4,RSIP4
ESRUE090
SWITCH, POLY (RAYCHEM) #RUE090
F1
ET2N3906
TRANSISTOR 2N3906 (MOT)
Q2,Q3
ETZVNL110A
TRANSISTOR, VOLTAGE REF.ZVNL110
Q1
EV25KKVSF8
20K POT
R1
EW1551B
WIRE 22AWG BLACK (ALPHA)
LCD1-16
EW1551R
WIRE 22AWG RED
LCD1-15
EXPX200
CERAMIC RESONATOR #PX200MC
X2
EXRN202-1.5
CHOKE (SCHAFFNER)
T1
EXSE3201
CRYSTAL 32.768 KHZ (EPSON)
X1
FCBSW11404
TERMINAL STRIP #BSW014-04-G-S
J5A
FCHTSW10912
TERMINAL STRIP (SAMTEC)
J2
COMPONENT LIST cont’d
TROUBLESHOOTING
SYMPTOM
REASON/CORRECTIVE ACTION
Inaccurate weight readings
Scale system MUST be zeroed with empty
wheelchair BEFORE patient is positioned in the
wheelchair and onto the scale (see directions for
BASIC SYSTEM OPERATION)
.
The characters only appear on half of the
display
Press the “
WEIGH
” button or remove one battery.
Wait five seconds, then re-install the battery and try
the “
WEIGH
” button again.
The display lights appear to work, but do
not respond to button activation
Membrane switches in the label may be damaged.
Make sure PENS or PENCILS are NOT used to press
buttons.
The display shows no reading at all
Check to ensure batteries are installed correctly (see
directions for
BATTERY REPLACEMENT)
.
Check display cable to make sure it is connected
securely.
For additional information or assistance, telephone the Service Hotline: 1-800-654-6360
or e-mail: