SR Instruments SRScales SR575 Скачать руководство пользователя страница 15

Model SR575 Bariatric Stand-On Scale 
Operating and Service Manual - S/N 1000+ 
Part No.  MAN575_031105 

 

Page 15 of 18 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

2003

 

S

Instruments, Inc. 

ER44991EJ

49.9K 1/8 WATT 1% J RESISTOR

R8

ER54991EJ

499K 1/8 WATT 1% J RESISTOR

R16

ER55491ED

549K 1/8 WATT 1% RESISTOR

R15

ER62004EJ

2M 1/8 WATT 1% J RESISTOR

R5,R14

ER72205QF

22M 1/4 WATT 5% RESISTOR

R9

ER770101

RESISTOR NETWORK 100K

RSIP1,2,3,4,RSIP4

ESRUE090

SWITCH, POLY (RAYCHEM) #RUE090

F1

ET2N3906

TRANSISTOR 2N3906 (MOT)

Q2,Q3

ETZVNL110A

TRANSISTOR, VOLTAGE REF.ZVNL110

Q1

EV25KKVSF8

20K POT

R1

EW1551B

WIRE 22AWG BLACK (ALPHA)

LCD1-16

EW1551R

WIRE 22AWG RED

LCD1-15

EXPX200

CERAMIC RESONATOR #PX200MC

X2

EXRN202-1.5

CHOKE (SCHAFFNER)

T1

EXSE3201

CRYSTAL 32.768 KHZ (EPSON)

X1

FCBSW11404

TERMINAL STRIP #BSW014-04-G-S

J5A

FCHTSW10912

TERMINAL STRIP (SAMTEC)

J2

COMPONENT LIST cont’d 

 

 

TROUBLESHOOTING 

 

SYMPTOM 

REASON/CORRECTIVE ACTION 

Inaccurate weight readings 

Scale system MUST be zeroed with empty 

wheelchair BEFORE patient is positioned in the 

wheelchair and onto the scale (see directions for 

BASIC SYSTEM OPERATION)

The characters only appear on half of the 
display 

Press the “

WEIGH

” button or remove one battery.  

Wait five seconds, then re-install the battery and try 

the “

WEIGH

” button again.   

The display lights appear to work, but do 
not respond to button activation 

Membrane switches in the label may be damaged.  

Make sure PENS or PENCILS are NOT used to press 

buttons.   

The display shows no reading at all 

Check to ensure batteries are installed correctly (see 

directions for 

BATTERY REPLACEMENT)

.  

Check display cable to make sure it is connected 

securely. 

For additional information or assistance, telephone the Service Hotline: 1-800-654-6360 

or e-mail: 

[email protected]

 

Содержание SRScales SR575

Страница 1: ...5_031105 Page 1 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com 2003 SInstruments Inc by SR Instruments In...

Страница 2: ...INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SYSTEM OPERATION 8 BATTERY REPLACEMENT 9 THEORY OF OPERATION 10 CALIBRATION 11 INITIALI...

Страница 3: ...truments com 2003 SInstruments Inc PACKING CHECKLIST Model SR575 Bariatric Stand On Scale DESCRIPTION QUANTITY BASE ASSEMBLY 1 ea MAST ASSEMBLY with DISPLAY and HEIGHT BAR 1 ea PACKAGE of SIX 6 D CELL...

Страница 4: ...ast Assembly and Base Assembly as indicated STEP 4 Gently rest the scale on its side Verify that the serial number on the bottom of the Display Unit 1 matches that on the Battery Compartment Cover 9 l...

Страница 5: ...d On Scale employs the latest in microprocessor and load cell technology to provide accurate and repeatable weight data Four 4 identically matched transducers are strategically placed to ensure an acc...

Страница 6: ...eral cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods longer than three 3 months remove the batteries To maintain proper operation of this instrumenta...

Страница 7: ...DOWN After 35 seconds HOLD While active freezes display data and stores it in memory LAST WEIGHT RECALL Hold button recalls last stored weight AUTO HOLD Locks display data and stores it in memory AVER...

Страница 8: ...viewed in either pounds or kilograms Pressing the LB KG MODE button allows the operator to toggle between the two readings Both pounds and kilograms are displayed in a resolution of 0 1 BASIC SYSTEM...

Страница 9: ...before zeroing the system press and hold the HOLD button for approximately five 5 seconds until the display reads AUTO HOLD ENABLED To return to CONTINUOUS WEIGH mode when finished press and hold the...

Страница 10: ...log to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to the refe...

Страница 11: ...known calibrated weight traceable to NIST onto the weighing surface and compare it to the displayed reading Note DO NOT USE barbell weights or calibrate to a mechanical scale STEP 4 Use the and butto...

Страница 12: ...ess buttons indicated to initialize the system The display will read HOLD TO INIT and count down from 9 to 0 When initialization is complete the display will read INITIALIZING and then return to the W...

Страница 13: ...g and Service Manual S N 1000 Part No MAN575_031105 Page 13 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments co...

Страница 14: ...C3 EC847RE016 470MFD RADIAL UVX1C471M C20 C21 ED1N5817 DIODE MOTOROLA D1 EICD4078BE IC CD4078BE TI IC4 EIDIP14 SOCKET 14 PIN DILB04P 223T IC4 EIDIP28 SOCKET 28 PIN DILB 28P 223T IC2 EIDIP40 SOCKET 40...

Страница 15: ...STAL 32 768 KHZ EPSON X1 FCBSW11404 TERMINAL STRIP BSW014 04 G S J5A FCHTSW10912 TERMINAL STRIP SAMTEC J2 COMPONENT LIST cont d TROUBLESHOOTING SYMPTOM REASON CORRECTIVE ACTION Inaccurate weight readi...

Страница 16: ...c s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc...

Страница 17: ...Operating and Service Manual S N 1000 Part No MAN575_031105 Page 17 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstru...

Страница 18: ...l S N 1000 Part No MAN575_031105 Page 18 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com 2003 SInstruments...