Model SR575 Bariatric Stand-On Scale 
Operating and Service Manual - S/N 1000+ 
Part No.  MAN575_031105 

 

Page 4 of 18 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

2003

 

S

Instruments, Inc. 

ASSEMBLY 

 

 

 

STEP 1

:  Unpack the scale system and 

check parts against the 

PACKING 

CHECKLIST

.  If there are any missing or 

damaged parts, please call the Service 

Hotline:  1-800-654-6360. 

STEP 2

: (Figure 1)

   

Position the Mast 

Assembly (2) into the Base Assembly (3) as 

shown, while feeding the display cable through 

the cable hole on the right.  Ensure the Height 

Bar (4) seats correctly in the Base Assembly. 

 

STEP 3: 

(Figure 1)  Insert the four (4) 3/8-16 

screws (5) into the holes in Mast Assembly and Base 

Assembly as indicated. 

STEP 4

:  Gently rest the scale on its side.  Verify 

that the serial number on the bottom of the Display 

Unit (1) matches that on the Battery Compartment 

Cover (9), located in the Base Assembly between the 

wheels.   

STEP 5

:  (Figure 2)  Open the Battery Compartment Cover.  Secure the Mast Assembly by 

screwing the four (4) 3/8”Nylon Insert Lock Nuts (6) onto the four (4) 3/8-16 Screws inserted in 

STEP 3. 

STEP 6

:  Remove the Height Bar Set Screws (7) and apply a small amount of Loctite

 

to each.  

Reinstall setscrews and tighten securely. 
 

PART NAME 

1

Display Unit 

2

Mast Assembly 

3

Base Assembly 

4

Height Bar 

5

3/8-16 Screws 

6

3/8” Nylon Insert Lock Nuts 

7

Height Bar Set Screws 

8

Display Cable Connector 

9

Battery Compartment Cover 

10

Leveling Feet  

Figure 1:  Scale Assembly Diagram 

Figure 2:  Display Cable Connector, Height Bar 

Set Screws, and Lock Nut Locations 

Cable 

Hole 

Continued next page

Содержание SRScales SR575

Страница 1: ...5_031105 Page 1 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com 2003 SInstruments Inc by SR Instruments In...

Страница 2: ...INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SYSTEM OPERATION 8 BATTERY REPLACEMENT 9 THEORY OF OPERATION 10 CALIBRATION 11 INITIALI...

Страница 3: ...truments com 2003 SInstruments Inc PACKING CHECKLIST Model SR575 Bariatric Stand On Scale DESCRIPTION QUANTITY BASE ASSEMBLY 1 ea MAST ASSEMBLY with DISPLAY and HEIGHT BAR 1 ea PACKAGE of SIX 6 D CELL...

Страница 4: ...ast Assembly and Base Assembly as indicated STEP 4 Gently rest the scale on its side Verify that the serial number on the bottom of the Display Unit 1 matches that on the Battery Compartment Cover 9 l...

Страница 5: ...d On Scale employs the latest in microprocessor and load cell technology to provide accurate and repeatable weight data Four 4 identically matched transducers are strategically placed to ensure an acc...

Страница 6: ...eral cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods longer than three 3 months remove the batteries To maintain proper operation of this instrumenta...

Страница 7: ...DOWN After 35 seconds HOLD While active freezes display data and stores it in memory LAST WEIGHT RECALL Hold button recalls last stored weight AUTO HOLD Locks display data and stores it in memory AVER...

Страница 8: ...viewed in either pounds or kilograms Pressing the LB KG MODE button allows the operator to toggle between the two readings Both pounds and kilograms are displayed in a resolution of 0 1 BASIC SYSTEM...

Страница 9: ...before zeroing the system press and hold the HOLD button for approximately five 5 seconds until the display reads AUTO HOLD ENABLED To return to CONTINUOUS WEIGH mode when finished press and hold the...

Страница 10: ...log to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to the refe...

Страница 11: ...known calibrated weight traceable to NIST onto the weighing surface and compare it to the displayed reading Note DO NOT USE barbell weights or calibrate to a mechanical scale STEP 4 Use the and butto...

Страница 12: ...ess buttons indicated to initialize the system The display will read HOLD TO INIT and count down from 9 to 0 When initialization is complete the display will read INITIALIZING and then return to the W...

Страница 13: ...g and Service Manual S N 1000 Part No MAN575_031105 Page 13 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments co...

Страница 14: ...C3 EC847RE016 470MFD RADIAL UVX1C471M C20 C21 ED1N5817 DIODE MOTOROLA D1 EICD4078BE IC CD4078BE TI IC4 EIDIP14 SOCKET 14 PIN DILB04P 223T IC4 EIDIP28 SOCKET 28 PIN DILB 28P 223T IC2 EIDIP40 SOCKET 40...

Страница 15: ...STAL 32 768 KHZ EPSON X1 FCBSW11404 TERMINAL STRIP BSW014 04 G S J5A FCHTSW10912 TERMINAL STRIP SAMTEC J2 COMPONENT LIST cont d TROUBLESHOOTING SYMPTOM REASON CORRECTIVE ACTION Inaccurate weight readi...

Страница 16: ...c s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc...

Страница 17: ...Operating and Service Manual S N 1000 Part No MAN575_031105 Page 17 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstru...

Страница 18: ...l S N 1000 Part No MAN575_031105 Page 18 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com 2003 SInstruments...

Отзывы: