Model SR575 Bariatric Stand-On Scale 
Operating and Service Manual - S/N 1000+ 
Part No.  MAN575_031105 

 

Page 10 of 18 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

2003

 

S

Instruments, Inc. 

THEORY OF OPERATION 

 
SR Instruments patient weighing systems are digital scales.  Strain-gauge force cells convert the force 
of an applied weight into an analog signal.  This signal is amplified by an operational amplifier and 
converted to a digital signal by an analog to digital converter.  The digital signal is transferred to a 
micro-controller where it is filtered, converted to appropriate units, and displayed on a liquid crystal 
display. 
Strain-gauge force cells each contain four strain gauges mounted in a full Wheatstone-bridge 
configuration.  These bridges convert the physical movement of the force cell, due to the applied mass 
on the system, into minute changes in electrical resistance.  These changes in resistance produce a 
voltage difference across the Wheatstone-bridge, which is amplified by the operational amplifier.  The 
amplifier is configured to current sum the output of each cell, with potentiometers serving to adjust the 
sensitivity (voltage out per unit of weight applied) of each bridge.  The offset potentiometer produces a 
small current, which nulls the output of the amplifier for an unloaded system. 
The output of the operational amplifier is digitized by the analog to digital converter.  The converter 
integrates the analog signal onto the integrating capacitor over a short interval.  The integrating 
capacitor is then discharged at a rate proportional to the reference voltage applied to the converter.  
The residual voltage on the integrating capacitor is then multiplied by a factor and again discharged at 
a rate proportional to the reference voltage.  The residual voltage from this discharge is again 
multiplied by a factor and again discharged.  The time taken to discharge the capacitor is proportional 
to the voltage from the operational amplifier, which is proportional to the applied load on the force 
cells.  The time is stored as a binary number in the analog to digital converter and is transferred to the 
micro-controller when the conversion is complete. 
The micro-controller averages and filters the digital output of the analog to digital converter, subtracts 
the value saved during the system zero operation and scales the filtered output, then displays the result 
on the liquid crystal display.  The micro-controller performs a rolling average of data for continuous 
weigh and, for AutoHold, the micro-controller averages the data before locking in on the reading. If 
the data variance is greater than 0.1% in the AutoHold mode, the micro-controller will reset the filter 
and start a new averaging period. 
The micro-controller can be placed in a calibration mode, where the system can be re-calibrated.  In 
the calibration mode, the result of the weigh operation is scaled to match the value by adjusting the 
“up” and “down” calibration buttons.  This new calibration factor is then stored in the non-volatile 
memory.

 

Содержание SRScales SR575

Страница 1: ...5_031105 Page 1 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com 2003 SInstruments Inc by SR Instruments In...

Страница 2: ...INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SYSTEM OPERATION 8 BATTERY REPLACEMENT 9 THEORY OF OPERATION 10 CALIBRATION 11 INITIALI...

Страница 3: ...truments com 2003 SInstruments Inc PACKING CHECKLIST Model SR575 Bariatric Stand On Scale DESCRIPTION QUANTITY BASE ASSEMBLY 1 ea MAST ASSEMBLY with DISPLAY and HEIGHT BAR 1 ea PACKAGE of SIX 6 D CELL...

Страница 4: ...ast Assembly and Base Assembly as indicated STEP 4 Gently rest the scale on its side Verify that the serial number on the bottom of the Display Unit 1 matches that on the Battery Compartment Cover 9 l...

Страница 5: ...d On Scale employs the latest in microprocessor and load cell technology to provide accurate and repeatable weight data Four 4 identically matched transducers are strategically placed to ensure an acc...

Страница 6: ...eral cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods longer than three 3 months remove the batteries To maintain proper operation of this instrumenta...

Страница 7: ...DOWN After 35 seconds HOLD While active freezes display data and stores it in memory LAST WEIGHT RECALL Hold button recalls last stored weight AUTO HOLD Locks display data and stores it in memory AVER...

Страница 8: ...viewed in either pounds or kilograms Pressing the LB KG MODE button allows the operator to toggle between the two readings Both pounds and kilograms are displayed in a resolution of 0 1 BASIC SYSTEM...

Страница 9: ...before zeroing the system press and hold the HOLD button for approximately five 5 seconds until the display reads AUTO HOLD ENABLED To return to CONTINUOUS WEIGH mode when finished press and hold the...

Страница 10: ...log to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to the refe...

Страница 11: ...known calibrated weight traceable to NIST onto the weighing surface and compare it to the displayed reading Note DO NOT USE barbell weights or calibrate to a mechanical scale STEP 4 Use the and butto...

Страница 12: ...ess buttons indicated to initialize the system The display will read HOLD TO INIT and count down from 9 to 0 When initialization is complete the display will read INITIALIZING and then return to the W...

Страница 13: ...g and Service Manual S N 1000 Part No MAN575_031105 Page 13 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments co...

Страница 14: ...C3 EC847RE016 470MFD RADIAL UVX1C471M C20 C21 ED1N5817 DIODE MOTOROLA D1 EICD4078BE IC CD4078BE TI IC4 EIDIP14 SOCKET 14 PIN DILB04P 223T IC4 EIDIP28 SOCKET 28 PIN DILB 28P 223T IC2 EIDIP40 SOCKET 40...

Страница 15: ...STAL 32 768 KHZ EPSON X1 FCBSW11404 TERMINAL STRIP BSW014 04 G S J5A FCHTSW10912 TERMINAL STRIP SAMTEC J2 COMPONENT LIST cont d TROUBLESHOOTING SYMPTOM REASON CORRECTIVE ACTION Inaccurate weight readi...

Страница 16: ...c s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc...

Страница 17: ...Operating and Service Manual S N 1000 Part No MAN575_031105 Page 17 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstru...

Страница 18: ...l S N 1000 Part No MAN575_031105 Page 18 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com 2003 SInstruments...

Отзывы: