2-01
DGP000814AAA
GSS6100 and SimCHAN user manual
© Spirent Communications Plc 2005 - 2007
6-5
Remote interface
6.2.6 EREF
Set External Reference Frequency
Sets the expected External Reference Frequency to the specified value. The
value is stored in non-volatile memory for use on power-up. The unit
automatically switches to external reference and seeks Phase Lock
whenever a signal is present on the rear panel connector.
The message format is:
EREF
1MHz | 5MHz | 10MHz
Note the parameter is an ASCII string and must be exactly as shown.
6.2.7 EREF
?
Query External Reference Frequency
Commands the unit to return an ASCII string describing the current External
Reference Frequency setting and the lock status.
The message format is:
EREF ?
Example Response:
EREF 10MHz INT
or
EREF 10MHz EXT UNLOCKED
or
EREF 10MHz EXT LOCKED
6.2.8 GPIB
Set the GPIB Primary Address
Set the IEEE-488 bus Primary Address of the device. The Primary Address
is saved in non-volatile memory. The unit is supplied with the IEEE-488
address set to 2. The change takes immediate effect and remains in force
indefinitely.
The message format is:
GPIB <gpib
address>
Where
<gpib address>
An integer in the range 1 to 30.