9 – Fault finding
OxyDetect Oxygen Deficiency Gas Detector
58
05311001B / Revision 4
Message
Measurement
screen icon
Recommended actions
Transducer error
Ensure that you are using the instrument in the
specified operating conditions (refer to Section 3).
If the fault persists, contact Servomex or your local
Servomex agent for assistance.
Tx incorrect type
Contact Servomex or your local Servomex agent
for assistance.
Tx Maintenance
Check that the sample gas concentration is not
higher than the transducer full scale range.
Recalibrate (both low and high) as described in
Section 7. If this does not clear the fault, turn the
instrument off, and then turn it on again. If the fault
message is then displayed again, contact
Servomex or your local Servomex agent for
assistance.
Tx not detected
Contact Servomex or your local Servomex agent
for assistance.
9.2
Viewing messages
9.2.1 Active messages
1.
Select
Status
Active.
2.
Each message status screen shows:
Date and time of message
The message type ("Fault", "Maintenance rqd" or "Service in Progress")
The message itself.
Figure 41 - The message status screen
3. Refer to Section 9.1 for the recommended actions associated with the displayed
messages.