S2/S9
4
Conditions on site
4.1
Ambient temperature
The effects of temperature on the zero signal and sensitivity are
compensated. For optimum measurement results, keep to the nominal
temperature range. Measurement errors due to temperature may occur if
there is uneven warming (e.g. by radiant heat) or cooling. Whilst a radiation
shield and all-round heat insulation provide noticeable improvements, they
should not be permitted to create any kind of force leakage.
4.2
Moisture
Extremes of humidity or tropical conditions are to be avoided where outside
the specified limit values (degree of protection IP65 to DIN EN 60529).
4.3
Contamination
Dust, dirt and other foreign bodies should not be allowed to accumulate to
such an extent that part of the measuring force can be diverted onto the
housing and falsify the measured value (force leakage).
NOTE:
The gap under the force introduction parts must not become
clogged with foreign bodies.
S2
S9
The overload trip
must be kept clear