Product Water - Low Production Rate
Cause
Corrective
Action
Plugged pre-filters
Replace pre-filters
Low water temperature
Heat feed water or use higher GPD membrane
Low feed pressure
Use booster pump or use higher GPD membrane
Fouled membrane
Replace membrane
The following chart illustrates the procedure for determination of RO membrane
performance. However, the chart represents only rough guidelines for
determining performance of RO membrane. Depending on your tap water
chemistry, the rejection characteristics of the membrane may vary significantly.
Method of Calculate %
Test Results
Conclusion
Testing
Rejection
TDS/
Measure feed water and Is Rejection
No - Replace Membrane
Conductivity RO product water TDS/
greater than
Yes - Membrane OK
Tester Conductivity
95%
?
Alkalinity
Measure feed water and Is Rejection
No - Replace Membrane
Test Kit
RO product water
greater than
Yes - Membrane OK
Alkalinity 90%
?
Hardness
*Measure feed water
Is Rejection
No - Replace Membrane
Test Kit
and RO product water
greater than
Yes - Membrane OK
Hardness 90%?
*Caution: This test is not to be used on softened water sources.
Temperature Extremes
1. The unit must be kept out of areas that are subject to freezing
temperatures.
2. High temperatures greater than 100° F (38° C) must be avoided. If
he unit is used outside, avoid putting the system in direct sunlight or
connecting it to a garden hose that may be exposed to sunlight.
SYSTEM TROUBLESHOOTING GUIDE: