11. Place the
yellow concentrate (waste) tubing
and the
blue purified (product)
water tubing
temporarily into a drain. Do not restrict flow from these lines.
12. Let both tubings run water down the drain for one hour.
13. Reinstall the DI cartridge into the DI housing. Hand tighten.
14. There is a group of red, white, and green sensor wires. At the end of the wires,
you will find a white connector. Just insert the connector into the receptacle
on the top of the TDS monitor. You can then use the Velcro on the back to
mount the meter wherever it’s convenient. For example, even on a wall for
easier viewing.
15. Open the cold water supply valve. The pressure should not exceed 80 psi.
16. Upon startup, air may be trapped in the DI cartridge (housing may not appear
full), this is a normal condition and it will not affect the operation of the DI
system. However, if you later install an ASO Valve/Check Valve/Float Valve for
automatic operation, all air must be kept purged from the housings by slightly
unscrewing the housing and letting the running water displace the air until the
housing is full of water. When full, hand tighten the housing.
17. Set the meter to “Line 3” and run the system until the meter reads zero. The
water is now ready to use.
You can set the meter to monitor either of the
3 probes at any time
. “Line 1” displays the Tap Water TDS. “Line 2” displays
the TDS of the RO Water. “Line 3” displays the TDS of the DI (product) Water.
18. Close the cold water supply valve.
19. Finish the installation by directing or connecting the yellow line to a
permanent drain.
20. Check the system to ensure that all fittings are tight and leak-free before leaving
the system unattended. (If anything is leaking, contact SpectraPure for assistance.)
NOTE: It may take several days of normal operation for the membrane to
reach its full production rate.
CHECK:
»
Ensure that all fittings are tight and leak-free before leaving the
system unattended.
»
The red concentrate (waste) line includes a smaller capillary tube
(flow restrictor) that is located “inside” of the tubing. Do not
remove or discard this restrictor - the system will not produce
permeate (product )water without the flow restrictor.
SYSTEM INSTALLATION & INITIALIZATION: (cont.)