DEIONIZATION CARTRIDGE REPLACEMENT:
(located in rightmost housing)
The DI stage is the SilicaBuster™ Extended Life SUPER DI™ Deionization
Cartridge, DI-SB-10HC. The condition of this cartridge should be judged by
observing the “Line 3” probe on the TDS monitor.
Turn the on the system and allow water to flow past the monitor probe for
at least 20 minutes before attempting to use the monitor. When the reading
on the TDS monitor (set to “Line 3”) displays “001”, it is time to replace the DI
cartridge.
Materials Required:
One DI-SB-10HC Deionization cartridge, filter wrench.
Procedure:
1. Remove the filter housing from its cap by rotating it to the left.
2. Remove and discard the old cartridge from the housing.
3. Thoroughly wash out the housings with hot soapy water to which a
few teaspoons of household bleach have been added. Rinse well with
clean hot water.
4. Install the new deionization cartridge. Make sure
the cartridge is installed in the correct direction as
marked on the filter housing and that the top seal
is securely attached to the top of the cartridge.
5. Reinstall the bottom housing onto the cap by
rotating it to the right and hand tighten only.
NOTE: Do not use filter wrench to tighten
housings. Over-tightening will damage
housings and void your warranty.
Warning:
SpectraPure does not recommend
drinking deionized water.