Troubleshooting: CVX-300 Fault Codes & Error Lamps
Fault
Code
Problem
Description
Try this…
1
No energy
detected at
detector head
Clean the fiber optic catheter tip and re‑calibrate at the
recommended settings (45/25 for ELCA devices; 60/40
for SLS devices.) Ensure the device is pointed at the
center of the energy detector, one to two inches away.
If calibration is unsuccessful, attempt to calibrate with
the Reference Catheter. If successful, try a new fiber
optic catheter.
If calibration of the Reference Catheter is unsuccessful,
record the Fault code and call Spectranetics
Customer Service.
2
3
4
Calibration
did not occur
within the
unit’s pre‑set
pulse limit
Beam
attenuator
at maximum
opening
Fiber Power
Monitor
test error
Clean the Energy Detector Surface with an alcohol prep.
Clean the fiber optic catheter tip and re‑calibrate at the
recommended settings (45/25 for ELCA devices; 60/40
for SLS devices.) Ensure the device is pointed at the
center of the energy detector, one to two inches away.
If calibration is unsuccessful, attempt to
re‑calibrate the device at a lower fluence
setting (40/25 for ELCA devices; 50/40 for SLS
devices.) If calibration is successful, proceed.
Report the incident to Spectranetics after the
case is complete.
If calibration is unsuccessful, turn the system OFF ( ),
then back ON ( | ). Attempt to calibrate with the
Reference Catheter. If successful, try a new fiber
optic catheter. If calibration of the Reference Catheter
is unsuccessful, record the Fault code and call
Spectranetics Customer Service.
5
Fiber optic
catheter handle
improperly
inserted
Remove the fiber optic catheter and reinsert. If the fault
is repeated, call Spectranetics Customer Service.
10‑50 System Error
detected
Record the Fault Code and call Spectranetics
Customer Service.