Excimer Laser System
Version 26
Operator’s Manual
English / English
9.
Depress and hold the Footswitch down until the laser stops and the “Cal OK’ light is
illuminated.
10.
Record the energy reading in mJ shown on the CVX-300 display panel.
11.
Depress the Ready button on the CVX-300 display panel.
12.
Aim the Reference Catheter directly at the center of the NIST Detector and Joule Meter.
13.
Depress the Footswitch and record the energy.
14.
Compare the recorded energy value in step 10 with the recorded energy value in step
13.
15.
The difference in the two recorded energy values should be less than 20% (CFR 21
1040.11 Section 1) when using the following equation
(energy value in step 10 – energy value in step 13 )
/ energy value in step 13
16.
Notify the Spectranetics Customer Service Department immediately if the difference in
the recorded energy values are greater than or equal to 20%.
17.
Press the Standby button, turn the key switch to the OFF (
) position, remove the
key and store it in a safe place, disconnect the Footswitch and store it in the front
compartment, disconnect the power cord from the power source and the laser, close
the catheter connector door, cover the laser system with the protective cover.