Troubleshooting
The keyswitch will not turn ON
( | )
the CVX‑300.
• Ensure the CVX‑300 power
cord is connected to the
proper source.
• Ensure the main circuit breaker,
located on the lower back
panel, is in the ON ( | ) position.
• Ensure the interlock plug is
inserted into the receptacle
located on the lower
back panel.
The buzzer sounds when
CVX‑300 power cord
is plugged in.
• Release the emergency
button by turning it in
a clockwise direction.
The CVX‑300 will not enter
the Calibrate mode.
• Allow the CVX‑300 time to
complete the warm‑up mode.
•
Insert a fiber optic catheter
into the coupler.
• Plug in the footswitch.
•
Set the fluence and pulse
repetition rate.
Alarm sounds when the ready
button is depressed.
• Calibrate the system.
The CVX‑300 does not
complete calibration.
•
Depress and hold
the
footswitch down until lasing
stops and the Cal OK indication
is illuminated.
Keyswitch is ON
( | )
but the
CVX‑300 will not come on after
the emergency button has
been reset.
• Turn the keyswitch to the
OFF (
) position and then
to the ON (
|
) position to reset
the system.
Fiber indicator is illuminated.
•
Ensure the fiber optic catheter
handle is fully inserted into
the coupler.
Service indicator is illuminated.
• Call Spectranetics Customer
Service for assistance.