![Spectranetics CVX-300 Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/spectranetics/cvx-300/cvx-300_operators-manual_1343921032.webp)
32
7030-0068-26
28Jul16
Customer Service, Canada/US (719) 633-8333 / (800) 231-0978
Customer Service, 31 33 434 7050
CVX-300
®
Excimer Laser System
Version 26
Operator’s Manual
English / English
Operational Modes
Setup Mode
1.
Remove the footswitch from the front storage compartment. Close the storage door
prior to operating the laser. Connect the footswitch plug into the receptacle located
on the rear panel.
2.
Connect the end of the power cord into the receptacle located on the lower back panel
of the CVX-300. Ensure power cord is fully seated into the power receptacle. Insert the
other end of the power cord into a wall receptacle with the proper output voltage.
Note: Anytime an extension cord is used to power the CVX-300 on a temporary basis,
it must be a minimum 12 AWG cord of type SJO, SJT, SJOOW or equivalent. The power
cord should be routed to avoid all tripping hazards and damage from other equipment.
Any temporary extension cord must be disconnected and removed immediately upon
completion of the task for which it was installed.
3.
Insert the key in the keyswitch on the control panel. Turn the keyswitch clockwise to
activate the system.
4.
When the system is activated, it enters Testing mode.
5.
Clean the surface of the energy detector with an alcohol prep before and after each
use.