![Spectranetics CVX-300 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/spectranetics/cvx-300/cvx-300_operators-manual_1343921018.webp)
18
7030-0068-26
28Jul16
Customer Service, Canada/US (719) 633-8333 / (800) 231-0978
Customer Service, 31 33 434 7050
CVX-300
®
Excimer Laser System
Version 26
Operator’s Manual
English / English
Stowable Energy Detector
The Energy Detector is located to the left of the laser/catheter coupler. The Energy Detector
is used to calibrate fiber optic catheters prior to use, and must be pulled up into position to
facilitate system calibration.
Return the Energy Detector to the storage position by pushing it back into the unit and
closing the protective cover.
NOTICE
At no time should the energy detector or its mounting pole be used in an attempt to move
the CVX-300.
As with all CVX-300 units, it is necessary to keep the area around the Catheter Connector,
Energy Detector, and Control Panel clean, dry and free of contaminants.
Open
Protective
Cover
Extend the Energy
Detector to the
fully up position
Push Energy
Detector back into
the unit and close
the Protective Cover