20
The Product Flow sensor has been calibrated at the factory during testing and isn
’
t
normally required during commissioning. If the product flow is not reading accurately,
confirm the product flow rate by following the Product Flow calibration steps.
Product Flow Calibration
4.
Allow the system to run for a few minutes to stabilize and then time in minutes and
seconds, how long it takes to fill a container of a known volume.
5.
Touch the
‘
Gal
’ (’
Liter
’)
field to enter the volume of the container used.
6.
Touch the
‘
Minutes
’
field to enter the minutes it took to fill the container.
Only enter
the minutes, ex: 3 min 15 sec should be entered as 3.
7.
Touch the seconds field to enter the seconds it took to fill the container.
Only enter
the seconds, ex: 3 min 15 sec should be entered as 15.
8.
Press
‘
Proceed
’.
You must save all changes when prompted after exiting the set-
tings menu
3. Press the Product Flow Button
1. Press the Menu Button
2. Press the Calibrate Sensors Button
4. Measure the product flow per the
process described below, enter the
numbers below and press Proceed.