BenchTop Cold Plate
The optional BenchTop cold plate can be cooled to -40 °C within 30 minutes when
fully insulated and covered. The cold plate will remain lower than -40 °C after the
cover is removed. These specifications are based on SP Industries test data from
similar units operating with no load.
Initial Setup of the BenchTop Cold Plate
Remove all packing materials carefully and inspect for shipping damage. In the event
that shipping damage has occurred, retain all packing material and immediately
contact your freight carrier.
PROPERTY CAUTION!
THE COLD PLATE IS CONNECTED TO THE BENCHTOP UNIT BY A FLEXIBLE
HOSE. THE FLEXIBLE HOSE IS CHARGED WITH REFRIGERANT AND THEREFORE SPECIAL CARE
SHOULD BE TAKEN WHEN REMOVING THE COLD PLATE AND BENCHTOP UNIT FROM THE SHIPPING
CONTAINERS.
THE FLEXIBLE HOSE MAY BE SHAPED OR FORMED WHEN THE UNIT IS AT ROOM
TEMPERATURE. DO NOT BEND THE HOSE WHILE AT LOW OPERATING TEMPERATURES
1.
Position the BenchTop unit on a stable, flat surface. The cold plate must be
fully insulated and have a removable cover (
i.e.
, 1/2-inch thick Styrofoam
board or polyurethane foam). This will ensure that the lowest possible
temperatures will be achieved. The hose between the plate and the unit is
very flexible and can also be inserted into the rear of the unit (approximately
10 inches) so that the length can be adjusted.
2.
Verify the outlet you intend to use provides the same voltage and ampacity
listed on the serial tag of your unit (
i.e.
, 100-115 Volts, 15 Amps, 60Hz).
3.
Verify the shelf thermocouple probe is inserted into the well at the rear of the
cold plate, and the thermocouple connector is plugged into the shelf probe
receptacle at the rear of the BenchTop cabinet.
Cold Plate Operation
1.
Ensure that the power cord is plugged in. Switch the circuit breaker on the
rear of the BenchTop cabinet to the “ON” position.
2.
Press the
COND
button on the front panel of the BenchTop to start the
compressor and begin cooling the cold plate. Wait approximately 30 minutes
for the condenser temperature to reach -40 °C. After about an hour, the unit
will be capable of achieving lower temperatures (
i.e.
, -50 °C).
3.
The cold plate is now ready for use.
Note
: When the cold plate is in operation and no product is being added, the cold plate
should be covered to prevent frost build-up.
4.
When finished, press the
COND
button to disable the refrigeration system and
disable cooling to the cold plate.
PRACTICAL OPERATING TIP.
THE ONLY BUTTON REQUIRED FOR OPERATING THE COLD PLATE IS
THE CONDENSER (
COND
) BUTTON. THE DEFROST, VACUUM AND AUTO BUTTONS DO NOT HAVE A
FUNCTION FOR THE COLD PLATE APPLICATION.
© SP Industries, Inc. 2009
Содержание VirTis BenchTop 2K
Страница 1: ...SP INDUSTRIES INC Freeze Dryer Operator s Manual VirTis BenchTop 2K 4K and 6K Freeze Dryers ...
Страница 2: ......
Страница 6: ...iv Revision 004 12 09 SP Industries Inc 2009 ...
Страница 10: ......
Страница 18: ...Installation and Setup VirTis BenchTop 2K 4K and 6K 8 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 30: ...Sentry 2 0 Controls VirTis BenchTop 2K 4K and 6K 20 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 36: ...Basic Operation VirTis BenchTop 2K 4K and 6K 26 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 44: ...Optional Components VirTis BenchTop 2K 4K and 6K 34 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 52: ...Appendix B Serum Vial Capacities VirTis BenchTop 2K 4K and 6K 42 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 54: ...Appendix C Stainless Steel Cleaners VirTis BenchTop 2K 4K and 6K 44 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 56: ...SP Industries 3538 Main Street Stone Ridge NY 12484 USA www SPIndustries com 800 824 0400 845 687 0071 ...