![SP VirTis BenchTop 2K Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/sp/virtis-benchtop-2k/virtis-benchtop-2k_operators-manual_1338598024.webp)
Sentry 2.0 Controls
VirTis BenchTop 2K, 4K and 6K
14
Revision 004, 12/09
© SP Industries, Inc. 2009
The
∧
symbol, located in the upper left corner of the screen, returns you to the
previous screen. The
>
symbol allows you to go to the next screen within a selected
option.
Vacuum Control
The
Vacuum Control field displays the current vacuum level control setpoint. To
change the setpoint, use the
Up/Down (+/-)
buttons and navigate to the Vacuum
Control field. Press and hold the
MENU
button. When prompted, use the
Up/Down
(+/-)
buttons to adjust the value. Press
MENU
to enter the new setpoint.
Purge (Optional)
The Purge function controls a VBS valve located between the vacuum pump and the
suction port of the unit. This allows high vacuum to be pulled within the volume
between the pump and the valve, without submitting the entire unit to high vacuum.
The gas ballast can then be opened and the oil cleansed within the pump.
Heat (Optional)
The Heat function gives the ability to add controlled heat to product shelves for
enhanced drying. If installed, you can set the temperature of the manifold shelves up
to 60 °C.
Note:
Optional features will only be displayed if the factory-installed options were purchased
with your unit.
User Options Menu
To access the User Options menu, highlight User Options from the Main Menu and
press the
MENU
button.
∧
User Options
Vacuum Control VLC
Temperature Units C
Vacuum Units Torr
Relay Test
>
Vacuum Pump Maintenance
>
Vacuum Control
The Vacuum Control field sets the method used for vacuum level control—None, VLC
or Power Save. Setting Vacuum Control to None disables vacuum level control.
Setting Vacuum Control to VLC allows gas to bleed into the system through a control
valve to maintain the vacuum setpoint. Setting Vacuum Control to Power Save
isolates the system and maintains pressure by shutting off power to the vacuum pump
and VBS valve. As the system pressure rises, the vacuum pump is re-energized and
the system is pulled back down. To avoid rapid cycling of the vacuum pump, the
Power Save Deadband should be set to 20 or more.
Содержание VirTis BenchTop 2K
Страница 1: ...SP INDUSTRIES INC Freeze Dryer Operator s Manual VirTis BenchTop 2K 4K and 6K Freeze Dryers ...
Страница 2: ......
Страница 6: ...iv Revision 004 12 09 SP Industries Inc 2009 ...
Страница 10: ......
Страница 18: ...Installation and Setup VirTis BenchTop 2K 4K and 6K 8 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 30: ...Sentry 2 0 Controls VirTis BenchTop 2K 4K and 6K 20 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 36: ...Basic Operation VirTis BenchTop 2K 4K and 6K 26 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 44: ...Optional Components VirTis BenchTop 2K 4K and 6K 34 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 52: ...Appendix B Serum Vial Capacities VirTis BenchTop 2K 4K and 6K 42 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 54: ...Appendix C Stainless Steel Cleaners VirTis BenchTop 2K 4K and 6K 44 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 56: ...SP Industries 3538 Main Street Stone Ridge NY 12484 USA www SPIndustries com 800 824 0400 845 687 0071 ...