![SP Scientific ULTRA-COOL Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/sp-scientific/ultra-cool/ultra-cool_operators-manual_1338670054.webp)
RS-232 Programmer’s Reference
Ultra-Cool
46
Rev 008, 01/12
© SP Scientific 2012
RS-232 Error Codes
The following error codes are returned “E # # #” if the parser/syntaxer/limits-checking
software finds fault with the received string.
Group 0: UART Transmit/Receive and General Receive errors.
E001 UART
Overrun.
Character received by UART before previous character has been retrieved
from the character input buffer. Microprocessor hardware failure.
E002 Parity
Error.
Character detected has incorrect parity as defined by control-panel
programming. See PARITY?
E003 Frame
Error.
Incorrect number of Stop Bits detected on received character. See
STOPBITS?
E004 Break
Error.
UART detects loss of RS-232 line with host. This error is only reported if such
a break occurs after the receipt of valid data, and before receipt of a <CR>
(carriage return). Line break is otherwise ignored.
E005
Line Too Long.
Received "line" (character string) exceeds 128 characters without receipt of
<CR> (carriage return).
E006
Receive Buffer Overrun.
1KB character input buffer is full. Commands are being received faster than
they can be processed.
E007
Line Buffer Overrun.
Lines are being received faster than they can be processed.
E008 Break
Error.
See E004.
Group 1 Errors: General Parsing and Transmission Errors
E010
Transmit Buffer Overflow.
Requests for information (queries) result in information being buffered faster
than it can be transmitted to the host (Baud-Rate dependent). 1KB buffer
exceeded.
E011
Parse Buffer Overflow.
Too many commands (>32) on a single line.
Содержание ULTRA-COOL
Страница 1: ...RECIRCULATING CHILLER OPERATOR S MANUAL ULTRA COOL WITH MPC CONTROL...
Страница 2: ......
Страница 6: ...iv Rev 008 01 12 SP Scientific 2012...
Страница 10: ...Introduction Ultra Cool 2 Rev 008 01 12 SP Scientific 2012...
Страница 26: ...System Control Ultra Cool 18 Rev 008 01 12 SP Scientific 2012...
Страница 34: ...System Control Ultra Cool 26 Rev 008 01 12 SP Scientific 2012...
Страница 38: ...Maintenance Ultra Cool 30 Rev 008 01 12 SP Scientific 2012...
Страница 58: ...RS 232 Programmer s Reference Ultra Cool 50 Rev 008 01 12 SP Scientific 2012...
Страница 63: ......