Ultra-Cool
Installation and Startup
Rev 008, 01/12
7
© SP Scientific 2012
External Connections
CAUTION: REMOTE CONNECTIONS SHOULD NOT BE RUN NEAR OR PARALLEL TO AC
POWER LINES OR IN THE VICINITY OF EQUIPMENT THAT GENERATES LARGE ELECTRICAL
FIELDS.
RS-232 Remote:
To comply with Semi-S2 requirements, the RS-232 cable should be shielded with
ferrites.
37-pin Remote I/O
To comply with Semi-S2 requirements, the 37-pin cable must be shielded.
Remote Temperature Sensor
An external 100-ohm RTD sensor is provided for remote control and temperature
display.
Remote EMO Override
Ensure the assembled Remote EMO override CPC connector, labeled RC-1462-A-1,
is inserted into the “Remote EMO” receptacle on the rear of the machine. If you do
not have the EMO Override CPC connector installed your machine WILL NOT
START.