System Control
Ultra-Cool
22
Rev 008, 01/12
© SP Scientific 2012
Display Decimals / Integer (dP = 1)
If set to one, decimals will be displayed, and decimal entry permitted. If set to zero,
decimal readout will be suppressed (integer format). Additionally, certain values (
e.g
.,
SP) will also be restricted to integer display and entry. However, residual decimals
will be carried internally. Should the user desire integer format, the SP should be
corrected to xx.0 prior to suppressing the decimal point.
Audible Alarm (nOIS = 1)
The audible alarm is a piezoelectric device and has three modes of operation.
Display Alarm
Applications
0
1
2
Off
5 second chirp Alarm
Continuous Sounding
Hospital
Normal Operation
Industrial Applications
Power Failure (PF = 0)
The power failure recovery mode setting determines what the controller will do when
power is restored following a power failure. The alphanumeric display will also show
“PF” which will indicate that there was a power failure at some time. The “PF” will be
displayed until any switch on the controller is pressed.
Temperature Scale (dEg = 0)
All temperature entries and readouts may be adjusted according to the following:
Display Scale
0 Celsius
1 Fahrenheit
2 Kelvin
Содержание ULTRA-COOL
Страница 1: ...RECIRCULATING CHILLER OPERATOR S MANUAL ULTRA COOL WITH MPC CONTROL...
Страница 2: ......
Страница 6: ...iv Rev 008 01 12 SP Scientific 2012...
Страница 10: ...Introduction Ultra Cool 2 Rev 008 01 12 SP Scientific 2012...
Страница 26: ...System Control Ultra Cool 18 Rev 008 01 12 SP Scientific 2012...
Страница 34: ...System Control Ultra Cool 26 Rev 008 01 12 SP Scientific 2012...
Страница 38: ...Maintenance Ultra Cool 30 Rev 008 01 12 SP Scientific 2012...
Страница 58: ...RS 232 Programmer s Reference Ultra Cool 50 Rev 008 01 12 SP Scientific 2012...
Страница 63: ......