Operating Your Lyophilizer
FTS Systems LyoStar™ II
20
Rev 012, 04/11
© SP Scientific 2011
After all of the parameters for the run have been entered, click on the Save As
2.
button to save the recipe for future use. To program a step, select a
temperature, ramp rate and hold time (also vacuum setpoint in the drying
phase).
To start the run, click on the Start button.
3.
The software will prompt, "Are you sure?", and present you with a checklist of
4.
items to double check before you begin the run.
Click Start, and the unit will begin executing the program you have chosen.
5.
The alphanumeric data collection window will open. Minimize this window at
6.
any time by touching the screen outside the data collection window.
Note: If you have not customized or deleted the alphanumeric data headings while
you were in the recipe screen, the headings that will print with the data will be from the
previous run.
If at any time during the run you require help programming or operating the
7.
system click on the Instructions button for assistance.
When the program has been fully executed press the Finished button on the
8.
Automatic or Data screens. If you would like to terminate the run before it is
fully executed press the Stop button on the Automatic or Data screens.
You will see a prompt asking if you want to back-fill or simply break vacuum to
9.
atmospheric. Follow the instructions on these screens making sure to stopper
prior to having the system return to atmospheric.
Note: Use the software to perform step 9. Do NOT use the condenser drain valve to
break vacuum. This sudden change to atmospheric pressure is not recommended for
the capacitance manometer.
You may view current data from various screens:
Содержание LYOSTAR II
Страница 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR II STANDARD...
Страница 2: ......
Страница 6: ...iv Rev 012 04 11 SP Scientific 2011...
Страница 29: ...FTS Systems LyoStar II Operating Your Lyophilizer Rev 012 04 11 21 SP Scientific 2011...
Страница 38: ...Data Management FTS Systems LyoStar II 30 Rev 012 04 11 SP Scientific 2011...
Страница 52: ...Optional Isolation Valve FTS Systems LyoStar II 44 Rev 012 04 11 SP Scientific 2011...
Страница 58: ...Optional Cleanroom Configuration FTS Systems LyoStar II 50 Rev 012 04 11 SP Scientific 2011...
Страница 66: ...General Maintenance FTS Systems LyoStar II 58 Rev 012 04 11 SP Scientific 2011...
Страница 70: ...Appendix A Troubleshooting FTS Systems LyoStar II 62 Rev 012 04 11 SP Scientific 2011...
Страница 71: ......