9.1. Measurement of the Polarization Index (PI) 

When pressing the 

Ε

 key during the tests, the Polarization Index (PI) value 

will be exhibited on the display. It is only possible to apply this function after 

a  minimum  of  10  minutes  of  measurement;  in  case  the  key  is  pressed 

before  this  minimum  limit,  the  display  will  show  the  message  of  value 

exhibition of PI value, but will not show any value. The polarization index is 

the quotient between the values of the insulation resistance measured both 

in 10 minutes and 1 minute. This index is useful to detect the damage of the 

insulation resistance by the excessive presence of dust, dirt and greases or 

through the action of chemical and physical agents. 

 

=

PI

R
R

10 minutes

1 minute

 

 

9.2. Measurement  of  the  Dielectric  Absorption  Index 
(DAI) 

When  pressing  the 

Ε

  key  during  the  tests,  the  Dielectric  Absorption  Index 

(DAI) value will be exhibited on the display. It is only possible to apply this 

function  after  a  minimum  of  1  minute  of  measurement;  in  case  the  key  is 

pressed  before  this  minimum  limit,  the  display  will  show  the  message  of 

value exhibition of DAI value, but will not show any value. The polarization 

index  is  the  quotient  between  the  values  of  the  insulation  resistance 

measured  at  the  60  and  30  second,  and  it  is  useful  for  preventive  and 

predictive  maintenance  of  windings  (present  in  transformers,  motors, 

generators, etc.) 

=

DAI

R
R

60 seconds

30 seconds

 

 

 

 

 

 

Содержание MD-5075 Series

Страница 1: ...3 MD 5075x 5kV Insulation tester User s Guide 2009 Sourcetronic All rights reserved...

Страница 2: ...tween the terminals while a measurement is running because it could be dangerous for the operator Be sure that there are not any voltage difference between the points to which the megohmmeter will be...

Страница 3: ...__________________________________15 9 1 Measurement of the Polarization Index PI _________________________16 9 2 Measurement of the Dielectric Absorption Index DAI _________________16 10 Other functi...

Страница 4: ...ctric absorption index automatic measurement Measurement of Leakage Current and Capacitance TIMER to program resistance test time Limit allowing making test type Pass Fail with programmable limit Step...

Страница 5: ...el control functions Voltage output terminal V Zero reference terminal R Guard G Terminal Display Keyboard High Voltage led On Off key RS 232 communication port Power supply input Printer Paper feed c...

Страница 6: ...dex DAI Backlight activates the display light Capacitance exhibit the capacitance value Indicates that the capacitance value is being exhibited Activated enables the programming of the Operation Mode...

Страница 7: ...is lower than programmed limit Start Start test Indicates that the test is being executed Stop End of test 4 Display Measurement results in the corresponding measuring unit elapsed time since the mea...

Страница 8: ...n of the MD 5075x till it is again turned off Note the battery losses part of its charge slowly even when it is not under use Before using this device or after not using it for a long time the battery...

Страница 9: ...in the picture are merely illustrative so there can be differences between them and the ones which really come with the equipment Observation At the time of the connection the equipment automatically...

Страница 10: ...measurement is performed one of the GUARD or R terminals must be grounded but not both of them simultaneously Technical Note 32 reproduced at the end of the manual explains the usage of GUARD termina...

Страница 11: ...e Test voltage is the only one parameter that may be modified during tests 8 2 Selection of the operation mode The MD 5075x megohmmeter has four operation modes Normal with TIMER SVT and Pass Fail The...

Страница 12: ...multiple of 500V up to a 5000V limit It is advisable to use the key in order to select this value and keys may be used but if the selected value is not a multiple of 500 it will be rounded down The t...

Страница 13: ...ing unit and the analogue indication will start by bar graph In order to end tests press the key At that moment last measured values will remain frozen in the MD 5075x display By pressing the key agai...

Страница 14: ...xcessive presence of dust dirt and greases or through the action of chemical and physical agents PI R R 10 minutes 1 minute 9 2 Measurement of the Dielectric Absorption Index DAI When pressing the key...

Страница 15: ...han 300M In these cases it is convenient to press the key before starting the measurement activating the filter which allows for the reaching of the insulation resistance value in an upward curve with...

Страница 16: ...the key without interrupting the tests When this key is pressed again the megohmmeter updates the resistance and chronometer values The led on key and the letter H on the display indicate that the fun...

Страница 17: ...elect the Hyper Terminal program In order to create a new connection enter a name select an icon and click on OK Choose the correct communication port in the next window After clicking on OK select th...

Страница 18: ...x will be printed after 1 minute and 10 minutes respectively Printing may be started or stopped at any time during the test However it is convenient to turn the printer on before starting the test in...

Страница 19: ...ls and connectors of the equipment must stay dry and clean Cleaning should be made using a wet cloth in water and a soft detergent or isopropyl alcohol be sure that the products to be used for cleanin...

Страница 20: ...cuit current Max 2mA Display Alphanumeric It exhibits measurements both digitally and analogically by bar graph Test voltages accuracy 3 of nominal value over a 10G resistance Megohmmeter accuracy 5 o...

Страница 21: ...IEC 61000 4 3 Electrostatic immunity In accordance with IEC 1000 4 2 Power supply Internal rechargeable battery NI MH 12 V 3 Ah mains of 100 240V Battery charger For 100 240V Operating temperature 5 C...

Страница 22: ...er The unknown resistance Rx is connected between Vt and R terminals Its value determines the current passing through the circuit which in turn is indicated by the microammeter The value of Rx can be...

Страница 23: ...duced during reading Through resistance R2 circulates a current which is not passing through the meter and consequently does not affect the reading In fact current through R2 originates a certain erro...

Страница 24: ...age in transit SOURCETRONIC will at its option repair or replace the defective equipment free of charge or refund your purchase price However if SOURCETRONIC determines that the failure was caused by...

Отзывы: