20
12. MegaLogg2 Software
This software makes communication between MD-5075x and a computer
with Windows operative system easier. It makes the transference of all the
data recorded in the apparatus memory, enabling the user to synchronize
the date and time of the equipment internal clock with the computer date
and time, to generate graphics and reports of tests and to clear the memory
of the megohmmeter. Installation and operation instructions are included in
the software.
13. Printer
Please, press
Α
key in order to enable the results automatic printing.
Measured values will be printed every 15 seconds, while the Dielectric
Absorption Index and Polarization Index will be printed after 1 minute and
10 minutes respectively. Printing may be started or stopped at any time
during the test. However, it is convenient to turn the printer on before
starting the test in order to print it completely, including the heading. After
last measurement, press PAPER FEED CONTROL key – placed in the
printer - three times to see the last lines before cutting the paper.
Attention: Do not pull the paper. Never try to put the paper back into the
printer. In any of these cases, the printer may be easily damaged
This MD-5075x printer uses 37mm-wide thermal paper, which comes in a
33mm-diameter reel. The figure shown below indicates how to introduce the
paper correctly.
White side
Yellow side
Press the PAPER FEED CONTROL key until the paper turns up. In order to
remove the old paper reel, cut the paper next to it and press the key PAPER
FEED CONTROL.