14

8.2.3. 

SVT Mode (step voltage tests) 

The use of 

Η

 key allows the MD-5075x setting for the performance of a step 

voltage  test;  when  this  mode  is  selected,  the  display  shows  the  SVT 

abbreviation. Under this operation mode, the user does not define a specific 

voltage  test,  but  a  maximum  voltage  value  the  device  will  start  tests 

applying a 500V voltage and increase this value in 500V steps each minute 

until  reaching  the  programmed  voltage.  At  each  stage,  the  MD-5075x 

measures the resistance before advancing towards the following step. 

 

The  use  of  voltage  adjusting  keys,  determines  the  value  of  the  highest 

voltage – which will be, in all cases, a multiple of 500V, up to a 5000V limit. 

It is advisable to use the 

Κ

 key in order to select this value; 

Ι

 and 

ϑ

  keys 

may  be  used,  but  if  the  selected  value  is  not  a  multiple  of  500,  it  will  be 

rounded  down.  The  test  result  is  calculated  according  to  the  following 

formula: 

 

=

SVT

R
R

V MAX

500

 

 

After test ending, the value may be recovered by pressing 

Ε

 

key. 

 

8.2.4. 

“Pass / Fail” Test mode 

Press 

Λ

 key in order to determine the lower insulation limit for type “Pass / 

Fail” test. Select this value using 

Θ

 and 

Ρ

 

keys. Possible values are 10M

100M

, 1G

Ω 

or 10G

 

During  a  “Pass/  Fail”  test,  the  MD-5075x  will  indicate  when  the  insulation 

resistance is lower than the programmed limit, both with an intermittent BIP 

and the 

Λ

 key led flashing . The 

Λ

 key led will remain flashing until the end 

of tests, or until the measurement of the resistance value is greater than the 

programmed limit. 

 

Содержание MD-5075 Series

Страница 1: ...3 MD 5075x 5kV Insulation tester User s Guide 2009 Sourcetronic All rights reserved...

Страница 2: ...tween the terminals while a measurement is running because it could be dangerous for the operator Be sure that there are not any voltage difference between the points to which the megohmmeter will be...

Страница 3: ...__________________________________15 9 1 Measurement of the Polarization Index PI _________________________16 9 2 Measurement of the Dielectric Absorption Index DAI _________________16 10 Other functi...

Страница 4: ...ctric absorption index automatic measurement Measurement of Leakage Current and Capacitance TIMER to program resistance test time Limit allowing making test type Pass Fail with programmable limit Step...

Страница 5: ...el control functions Voltage output terminal V Zero reference terminal R Guard G Terminal Display Keyboard High Voltage led On Off key RS 232 communication port Power supply input Printer Paper feed c...

Страница 6: ...dex DAI Backlight activates the display light Capacitance exhibit the capacitance value Indicates that the capacitance value is being exhibited Activated enables the programming of the Operation Mode...

Страница 7: ...is lower than programmed limit Start Start test Indicates that the test is being executed Stop End of test 4 Display Measurement results in the corresponding measuring unit elapsed time since the mea...

Страница 8: ...n of the MD 5075x till it is again turned off Note the battery losses part of its charge slowly even when it is not under use Before using this device or after not using it for a long time the battery...

Страница 9: ...in the picture are merely illustrative so there can be differences between them and the ones which really come with the equipment Observation At the time of the connection the equipment automatically...

Страница 10: ...measurement is performed one of the GUARD or R terminals must be grounded but not both of them simultaneously Technical Note 32 reproduced at the end of the manual explains the usage of GUARD termina...

Страница 11: ...e Test voltage is the only one parameter that may be modified during tests 8 2 Selection of the operation mode The MD 5075x megohmmeter has four operation modes Normal with TIMER SVT and Pass Fail The...

Страница 12: ...multiple of 500V up to a 5000V limit It is advisable to use the key in order to select this value and keys may be used but if the selected value is not a multiple of 500 it will be rounded down The t...

Страница 13: ...ing unit and the analogue indication will start by bar graph In order to end tests press the key At that moment last measured values will remain frozen in the MD 5075x display By pressing the key agai...

Страница 14: ...xcessive presence of dust dirt and greases or through the action of chemical and physical agents PI R R 10 minutes 1 minute 9 2 Measurement of the Dielectric Absorption Index DAI When pressing the key...

Страница 15: ...han 300M In these cases it is convenient to press the key before starting the measurement activating the filter which allows for the reaching of the insulation resistance value in an upward curve with...

Страница 16: ...the key without interrupting the tests When this key is pressed again the megohmmeter updates the resistance and chronometer values The led on key and the letter H on the display indicate that the fun...

Страница 17: ...elect the Hyper Terminal program In order to create a new connection enter a name select an icon and click on OK Choose the correct communication port in the next window After clicking on OK select th...

Страница 18: ...x will be printed after 1 minute and 10 minutes respectively Printing may be started or stopped at any time during the test However it is convenient to turn the printer on before starting the test in...

Страница 19: ...ls and connectors of the equipment must stay dry and clean Cleaning should be made using a wet cloth in water and a soft detergent or isopropyl alcohol be sure that the products to be used for cleanin...

Страница 20: ...cuit current Max 2mA Display Alphanumeric It exhibits measurements both digitally and analogically by bar graph Test voltages accuracy 3 of nominal value over a 10G resistance Megohmmeter accuracy 5 o...

Страница 21: ...IEC 61000 4 3 Electrostatic immunity In accordance with IEC 1000 4 2 Power supply Internal rechargeable battery NI MH 12 V 3 Ah mains of 100 240V Battery charger For 100 240V Operating temperature 5 C...

Страница 22: ...er The unknown resistance Rx is connected between Vt and R terminals Its value determines the current passing through the circuit which in turn is indicated by the microammeter The value of Rx can be...

Страница 23: ...duced during reading Through resistance R2 circulates a current which is not passing through the meter and consequently does not affect the reading In fact current through R2 originates a certain erro...

Страница 24: ...age in transit SOURCETRONIC will at its option repair or replace the defective equipment free of charge or refund your purchase price However if SOURCETRONIC determines that the failure was caused by...

Отзывы: