20. PROGRAMMABLE PARAMETERS
20.2.
VENTRICULAR TACHYARRHYTHMIA DETECTION
Therapy zones
Values
Slow VT detection zone
(1)
Slow VT ON-
Slow VT OFF
VT detection zone
VT ON-
VT OFF
Fast VT / VF detection zone
Fast VT+VF ON-
VF ON
Slow VT rate (lower limit) (bpm)
From 100 to 200 by steps of 5 ;
190
VT rate (lower limit) (bpm)
130-135-140-145-150-155-160-165-170-175-180-185-
190
-
195-200-210-220-230
VF rate (lower limit) (bpm)
150-155-160-165-170-175-180-185-
190
-195-200-210-220-
230-240
Fast VT rate (upper limit) (bpm)
155-160-165-170-175-180-185-
190
-195-200-210-220-230-
240-255
Slow VT persistence (cycles)
4-6-8-
12
-16-20-30-50-100-200
VT persistence (cycles)
4-6-8-
12
-16-20-30-50-100-200
VF persistence (cycles)
From 4 to 20 by steps of 1 ;
6
(1) The Slow VT zone should be programmed ON only if the VT zone is programmed ON.
Detection criteria
Values
Slow VT and VT detection criteria
Rate Only-Stability-St-Stability/Acc-St/Acc-
PARAD
-PARAD+
Fast VT detection criteria
Rate+Stability
-Rate Only
Majority: (X/Y), Y (cycles)
8
-12-16
Majority: (X/Y), X (%)
65-70-
75
-80-90-95-100
Window of RR stability for Slow VT
and VT (ms)
30-45-
65
-80-95-110-125
Window of RR stability for fast VT
(ms)
30
-45-65
Acceleration (%)
6-13-19-
25
-31-38-44-50
Long cycle persistence extension
(cycles)
From 0 to 16 by steps of 1 ;
10
Long cycle gap (ms)
15-30-45-65-80-95-110-125-140-155-
170
-190-205
Atrial monitoring
Yes
-No
SORIN – INTENSIA DR 154 – U152A
53