20. PROGRAMMABLE PARAMETERS
Shock program
Values
Shock 1 (J)
OFF
-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-
10-12-14-16-18-20-22-24-26-28-30-32-34-42
Shock 2 (J)
OFF
-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-
10-12-14-16-18-20-22-24-26-28-30-32-34-42
Number of Max. Shock (42 J)
1-2-3-
4
20.4.
REMOTE ALERTS AND WARNINGS
The device routinely performs security self-checks and technical measurements to ensure
system integrity. When system integrity is found to be at risk outside a follow-up, alerts are
stored in the device memory. When system integrity is found to be at risk during a follow-up,
the information is managed as a warning (pop-up message) to notify immediately the user.
For example, the following types of event can trigger a warning or an alert: technical
problem during a shock, pacing lead impedance or coil impedance measurements out-of-
range, battery depletion, etc. The Remote tab presents an overview of all the alerts
managed by the device.
General parameters
Values
RF communication
(1)
ON-
OFF
Remote alerts
(1)
ON
-OFF
(1) RF and Remote alerts are turned on automatically if Shocks are programmed ON.
When Alerts are programmed "On", the following System Alerts are automatically activated:
─
Battery depletion – RRT
─
Device reset
─
Excessive charge time (>25s)
─
System integrity
Lead Alerts
Values
Abnormal A lead impedance
ON
-OFF
Abnormal A lead low limit (Ohm)
200
-250-300-350-400-450-500
Abnormal A lead high limit (Ohm)
1500-1750-2000-2500-
3000
Abnormal V lead impedance
ON
-OFF
Abnormal V lead low limit (Ohm)
200
-250-300-350-400-450-500
Abnormal V lead high limit (Ohm)
1500-1750-2000-2500-
3000
Abnormal RV coil impedance
ON
-OFF
Abnormal SVC coil impedance
ON
-OFF
Abnormal Shock impedance
(1)
ON
-OFF
(1) Normal impedance range [20 Ohm-200 Ohm]
SORIN – INTENSIA DR 154 – U152A
57