16. MAIN FUNCTIONS
─
battery status
─
lead status (brady leads and defibrillation coils)
─
pacing counters and mean heart rate (brady)
─
atrial and ventricular arrhythmia counters and episodes
─
ventricular therapy counters
Data are presented in the form of 2 reports to the physician: the first one contains a
summary of major counters, histograms, warnings and diagnosis. The second one presents
the 3 most important IEGM episodes automatically selected based on the degree of severity
for the patient.
16.7.4 .
User website
On the website, the physician is able to:
─
consult and schedule the remote follow-ups of their patient
─
configure additional ways of being notified of alerts (for instance by SMS, fax or e-mail
─
consult, print and export patient reports
16.7.5 .
Alert system
The following set of alert trigger can be independently programmed ON/OFF by the
physician using the dedicated programmer and can trigger an alert transmission:
─
Low or high impedance (A, V)
─
Abnormal coil impedance (shock lead)
─
Low or High shock impedance
─
Inefficient high energy shock
─
All shocks programmed OFF
─
Shock treated VT/VF
─
Suspicion of noise on the V lead
─
AF occurrence
─
Fast V rate during AF
SORIN – INTENSIA DR 154 – U152A
43
WARNING: The use of remote monitoring does not replace regular follow-up. Therefore,
when using remote monitoring, the time period between follow-ups visits may not be
extended.