1-21
UP-55MD
1-9. Lithium Battery
The KY-582 board is mounted a lithium battery for the clock.
The lithium battery is not available as repair parts. When replacing, use a CR2025-type lithium coin
battery by Sony.
1.
Open the front door.
2.
Remove the battery cover using a fine-tipped driver, etc..
3.
Remove the lithium battery in the direction of the arrow from battery holder on the KY-582 board.
4.
To install, reverse the removal procedure.
n
When attaching the lithium battery, attach with the
+
side facing upward.
Lithium battery
Front door
Battery cover
Lithium battery
KY-582 board
Place the
+
side
facing upward.
KY-582 board