Performance Tests
AWG2021 Service Manual
4-29
Electrical
Characteristic
Checked
Arithmetic Operation, Add, on pageĂ1-13.
Equipment
Required
Two 50Ă
W
coaxial cables
One function generator
One digital multimeter (DMM)
Prerequisites
The AWG2021 meets the prerequisites listed on page 4-7. Option 02
(CH2 output channel) must be installed in the AWG2021.
1. Use the test hookup and control settings from the previous check.
2. Check internal Add operation:
a. Check Autostep Step 4:
H
Push the AWG2021 MANUAL TRIGGER button, and check that
the step changes to Step: 4 on the MODE menu.
H
Check that the DMM reading is in the range from +2.85 V to
+3.15 V.
3. End procedure:
a. Disable function generator output: Turn off the function generator
output.
b. Remove equipment: Disconnect connections to the test equipment.
Check Internal Add
Operation (Option 02
Only)
Содержание Tektronix AWG2021
Страница 4: ......
Страница 16: ...Service Safety Summary xii AWG2021 Service Manual ...
Страница 21: ...Specifications ...
Страница 22: ......
Страница 39: ...Operating Information ...
Страница 40: ......
Страница 60: ...Operating Instructions 2 20 AWG2021 Service Manual ...
Страница 61: ...Theory of Operation ...
Страница 62: ......
Страница 66: ...Theory of Operation 3 4 AWG2021 Service Manual ...
Страница 67: ...Performance Verification ...
Страница 68: ......
Страница 123: ...Adjustment Procedures ...
Страница 124: ......
Страница 135: ...Maintenance ...
Страница 136: ......
Страница 144: ...Maintenance 6 8 AWG2021 Service Manual ...
Страница 188: ...Removal and Installation Procedures 6 52 AWG2021 Service Manual ...
Страница 196: ...Troubleshooting 6 60 AWG2021 Service Manual TP100 5V TP101 GND Figure 6 34 Key Board ...
Страница 201: ...Options ...
Страница 202: ......
Страница 222: ...Options and Accessories 7 20 AWG2021 Service Manual ...
Страница 223: ...Replaceable Electrical Parts ...
Страница 224: ......
Страница 226: ...Replaceable Electrical Parts 8 2 AWG2021 Service Manual ...
Страница 227: ...Diagrams ...
Страница 228: ......
Страница 230: ...Diagrams 9 2 AWG2021 Service Manual ...
Страница 236: ...9 8 AWG2021 Service Manual ...
Страница 237: ...Replaceable Mechanical Parts ...
Страница 238: ......
Страница 253: ...Replaceable Parts AWG2021 Service Manual 10 15 4 5 1 2 3 12 13 11 1 11 10 A23 A31 6 7 8 9 Figure 10 6 Option 02 ...
Страница 255: ...Replaceable Parts AWG2021 Service Manual 10 17 2 5 1 3 4 A9 2 Figure 10 7 Option 03 ...
Страница 257: ...Replaceable Parts AWG2021 Service Manual 10 19 3 1 2 A25 4 Figure 10 8 Option 04 ...
Страница 261: ......
Страница 262: ......