NOTE
. For operation to specified accuracy, allow the AWG2021 to warm up at
least 20 minutes before doing the performance tests.
H
Load all the files from the Performance Check/Adjustment disk
(063-2171-XX) that comes with this manual into AWG2021 internal
memory. For instructions on loading files, see Loading Files on page 2–14 in
the Instructions for Operation subsection in section 2.
Read Preparation and Conventions on page 4-1. Also, if you are not familiar
with operating the AWG2021, read the subsection Instructions for Operation, in
section 2 before doing any of these procedures.
The following equipment is required to check the performance of the AWG2021.
Table 4-1: Test Equipment
Item Description
Minimum Requirements
Example
Purpose
Precision termination
Impedance: 50Ă
W
, 0.1%
Connectors: BNC
Tektronix Part 011Ć0129ĆXX
Signal termination.
Adapter
Connectors: BNC femaleĆtoĆdual
banana
Tektronix Part 103Ć0090ĆXX
Signal interconnection.
Adapter
Connectors: SMA maleĆtoĆBNC
female
Tektronix Part 015Ć0554ĆXX
Signal interconnection.
BNC dual input (TEE)
adapter
Connectors: BNC
Tektronix Part 103Ć0030ĆXX
Signal interconnection.
BNC cable
(4 required)
Impedance: 50Ă
W
Connectors: BNC
Length: 43 inches
Tektronix Part 012Ć0057ĆXX
Signal interconnection.
Termination board
(Option 03)
Must use example equipment
Tektronix Part 671Ć2957ĆXX
Used to check ECL digital
data output.
ECL Digital Data Out
Cable (Option 03)
Must use example equipment
Tektronix Part 012Ć1408ĆXX
(Standard accessory for Option 03)
Used to check ECL digital
data output.
Test leads
(Option 03)
Must use example equipment
Tektronix Part 012Ć1381ĆXX
Used to check ECL digital
data output.
TTL Digital Data Out
Cable (Option 04)
Must use example equipment
Tektronix Part 174Ć3192ĆXX
(Standard accessory for Option 04)
Used to check TTL digital
data output.
2
×
13 header
(Option 04)
Must use example equipment
Tektronix Part 131Ć3847ĆXX
Used to check TTL digital
data output.
Probe, 10X
(Option 04)
10X probe
Tektronix Part P6139A
Used to check TTL digital
data output.
Test oscilloscope
Bandwidth: >250ĂMHz
Tektronix TDS500 Series Digitizing
Oscilloscope or 2400 Series DigitizĆ
ing Oscilloscope
Checks output signals.
Used in many procedures.
Related Information
Equipment Required
Содержание Tektronix AWG2021
Страница 4: ......
Страница 16: ...Service Safety Summary xii AWG2021 Service Manual ...
Страница 21: ...Specifications ...
Страница 22: ......
Страница 39: ...Operating Information ...
Страница 40: ......
Страница 60: ...Operating Instructions 2 20 AWG2021 Service Manual ...
Страница 61: ...Theory of Operation ...
Страница 62: ......
Страница 66: ...Theory of Operation 3 4 AWG2021 Service Manual ...
Страница 67: ...Performance Verification ...
Страница 68: ......
Страница 123: ...Adjustment Procedures ...
Страница 124: ......
Страница 135: ...Maintenance ...
Страница 136: ......
Страница 144: ...Maintenance 6 8 AWG2021 Service Manual ...
Страница 188: ...Removal and Installation Procedures 6 52 AWG2021 Service Manual ...
Страница 196: ...Troubleshooting 6 60 AWG2021 Service Manual TP100 5V TP101 GND Figure 6 34 Key Board ...
Страница 201: ...Options ...
Страница 202: ......
Страница 222: ...Options and Accessories 7 20 AWG2021 Service Manual ...
Страница 223: ...Replaceable Electrical Parts ...
Страница 224: ......
Страница 226: ...Replaceable Electrical Parts 8 2 AWG2021 Service Manual ...
Страница 227: ...Diagrams ...
Страница 228: ......
Страница 230: ...Diagrams 9 2 AWG2021 Service Manual ...
Страница 236: ...9 8 AWG2021 Service Manual ...
Страница 237: ...Replaceable Mechanical Parts ...
Страница 238: ......
Страница 253: ...Replaceable Parts AWG2021 Service Manual 10 15 4 5 1 2 3 12 13 11 1 11 10 A23 A31 6 7 8 9 Figure 10 6 Option 02 ...
Страница 255: ...Replaceable Parts AWG2021 Service Manual 10 17 2 5 1 3 4 A9 2 Figure 10 7 Option 03 ...
Страница 257: ...Replaceable Parts AWG2021 Service Manual 10 19 3 1 2 A25 4 Figure 10 8 Option 04 ...
Страница 261: ......
Страница 262: ......