5-28
SEL1655G
4. Analysis of each image file is automatically started.
[][][][]
[][][][][][][][][][][]
5. When the analysis is completed, “<TELE> OK” is displayed on
the message.
[][][][]
[][][][][][][][][][][]
6. Click the [WIDE START] button on the RACS screen and con-
firm that the “<WIDE> Drag-and-drop image file(s) here.” is
displayed on the message.
[][][][]
[][][][][][][][][][][]
7. From the 42 image files captured in the PC, select the 21 image
files shot with WIDE, and drag & drop to the RACS screen.
[][][][]
[][][][][][][][][][][]
8. Analysis of each image file is automatically started.
[][][][]
[][][][][][][][][][][]
9. When the analysis is completed, “<WIDE> OK” is displayed on
the message.
[][][][]
[][][][][][][][][][][]
When “SEL1655G_1ST_LENS_ASSY” is selected:
t
Proceed to step 10
When “SEL1655G_MOUNT_ASSY” is selected:
t
Proceed to step 16
10. Click the [START SHIM CALCULATION] button and the
ShimCaluc screen is appeared. Input the recorded current washer
thickness in the current area.
SEL1655G_1ST_LENS_ASSY
[][][][]
Current
area
– Continued on next page –