Important EMC notices for use in the medical environments
• The PDW-75MD needs special precautions regarding EMC
and needs to be installed and put into service according to
the EMC information provided in this instructions for use.
• The portable and mobile RF communications equipment
such as cellular phones can affect the PDW-75MD.
Warning
The use of accessories and cables other than those specified,
with the exception of replacement parts sold by Sony
Corporation, may result in increased emissions or decreased
immunity of the PDW-75MD.
Warning
If the PDW-75MD should be used adjacent to or stacked with
other equipment, it should be observed to verify normal
operation in the configuration in which it will be used.
List of cables used for EMC tests
Type of cable
Specifications
AC mains cable
2.4 m, non-shielded
75-ohm BNC coaxial cable
1.5 m, shielded
XLR 3-pin Cannon cable
5 m, shielded
RCC-5G cable
3.2 m, shielded
Analog RGB cable
2 m, shielded
RCA pin / BNC cable
5 m, shielded
XLR 3-pin / RCA pin cable
5 m, shielded
i.LINK cable
1.5 m, shielded
Guidance and manufacturer’s declaration - electromagnetic emissions
The PDW-75MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the PDW-75MD should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The PDW-75MD uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The PDW-75MD is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies