32
Menu Operation
Adj
u
s
tme
nt
Chart of the high tone range
2
Display the selected number by pressing the
F
or
f
button.
To reset to the default setting of the high tone
range
Display 75.
13
Display the “Save Gamma” by pressing the EXEC
button.
14
Press the EXEC button.
The gamma curve adjustment values selected in step
4 are stored and will be effective from the next
printing operation.
The following screen appears on the LCD panel.
Press the MENU button to close the menu screen and
return to the standard screen.
Calibrating the Printing Density
The unit automatically calibrates the printing density
according to the density curve set.
Perform calibration for each type of blue thermal film
separately.
It is recommended to perform this operation when new
film was inserted.
1
Press the MENU button to bring up the menu screen.
The next darkest part
The darkest part
The tone of
the border of
each box on
the chart can
be changed to
that of the
borders
above or
below it on
the chart.
The tone of the center of each box on the chart can
be changed to that of the boxes to the right or left of
it as shown in the chart.
1
2
3
3