Chapter 6: Safety
Biological Safety
Electromagnetic Compatibility (EMC)
The ultrasound system has been tested and found to comply with the electromagnetic compatibility
(EMC) limits for medical devices to IEC 60601-1-2:2001. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation.
Non AP/APG
Ultrasound system power supply, Mobile Docking
System, and peripherals. Equipment is not suitable for
use in the presence of flammable anaesthetics.
WARNING:
To avoid the risk of a burn hazard, do not use the transducer with high frequency
surgical equipment. Such a hazard may occur in the event of a defect in the high
frequency surgical neutral electrode connection.
Caution:
Medical electrical equipment requires special precautions regarding EMC and must
be installed and operated according to these instructions. It is possible that high
levels of radiated or conducted radio-frequency electromagnetic interference (EMI)
from portable and mobile RF communications equipment or other strong or nearby
radio-frequency sources, could result in performance disruption of the ultrasound
system. Evidence of disruption may include image degradation or distortion, erratic
readings, equipment ceasing to operate, or other incorrect functioning. If this
occurs, survey the site to determine the source of disruption, and take the following
actions to eliminate the source(s).
• Turn equipment in the vicinity off and on to isolate disruptive equipment.
• Relocate or re-orient interfering equipment.
• Increase distance between interfering equipment and your ultrasound system.
• Manage use of frequencies close to ultrasound system frequencies.
• Remove devices that are highly susceptible to EMI.
• Lower power from internal sources within facility control (such as paging systems).
• Label devices susceptible to EMI.
• Educate clinical staff to recognize potential EMI-related problems.
• Eliminate or reduce EMI with technical solutions (such as shielding).
• Restrict use of personal communicators (cell phones, computers) in areas with
devices susceptible to EMI.
• Share relevant EMI information with others, particularly when evaluating new
equipment purchases which may generate EMI.
• Purchase medical devices that comply with IEC 60601-1-2 EMC Standards.
Содержание titan
Страница 1: ...TITAN Ultrasound System User Guide Supplement ...
Страница 2: ......
Страница 4: ......
Страница 5: ...简体中文 TITAN 超声仪用户指南 补充说明 ...
Страница 18: ...12 Chapter 6 Safety ...
Страница 19: ...Ergänzung zum TITAN Ultraschallsystem Benutzerhandbuch ...
Страница 35: ...Suplemento del Manual para el usuario del sistema de ecografía TITAN ...
Страница 51: ...Supplément au Guide d utilisation de l échographe TITAN ...
Страница 66: ...14 Chapitre 6 Sécurité ...
Страница 67: ...Supplemento al Manuale dell utente del sistema per ecografia TITAN ...
Страница 82: ...14 Capitolo 6 Sicurezza ...
Страница 83: ...Suplemento do Manual do Usuário do Sistema de Ultra som TITAN ...
Страница 98: ...14 Capítulo 6 Segurança ...
Страница 99: ...Tillæg til brugervejledning til TITAN ultralydssystem ...
Страница 113: ...Tillägg till användarhandbok för ultraljudssystemet TITAN ...
Страница 128: ...14 Kapitel 6 Säkerhet ...
Страница 129: ...TITAN 超声仪用户指南 补充说明 ...
Страница 142: ...12 第 6 章 安全性 ...
Страница 143: ......
Страница 144: ...P06816 01 ...