● S-80 DC ● S-110 DC ● S-120 DC – USER MANUAL 

2.2 

Functionality 

 

Before  the  set  can  be  used,  the  operating  unit  and  high-voltage  unit  have  to  be  grounded 
properly. 

 

When  the  high-voltage  adjuster  is  set  to  zero,  the  test  voltage  can  be  switched  on  at  the 
press  of  a  button  and  increased  continuously  between  0  V  and  the  nominal  voltage.  This 
voltage is generated in the high-voltage unit using transformers with rectifiers connected in 
outgoing circuit. 

 

The condition of the test object with regard to electric strength and insulation quality can be 
determined from the values for test voltage and leakage current displayed on the  operating 
unit. 

 

When testing has been completed, the test object is discharged and short-circuited with the 
grounding and discharge rod. 

2.3 

Application 

The  S-50  /  80  /  110  /  120  DC  is  used  for  electrical  testing  of  power  cables  and  plant  with  direct 

voltage. Areas of application include: 

 

testing at re-commissioning of cable systems after maintenance and repair, 

 

testing of newly installed cables an cable joints before commissioning, 

 

voltage testing at cable sheath, 

 

regular cable testing, 

 

testing of electrical equipment. 

The following test voltages are recommended by HD 621 S1 for testing with direct voltage: 
 

Nominal voltage U

0

/U [kV/kV] 

3.6/6 

6/10 

12/20 

18/30 

20.8/36 

DC test voltage [kV] 

20…29 

34…48 

67…96 

76…108 

87…124 

 

 

WARNING 
 

 

Owing to the high voltages which occur, special safety measures are required.  
These are explained in later sections of the manual (see also sec. 1).
 

 

  The  digital  meters  only  work  when  the  tester  is  powered  on.  In  the 

case of power loss check with external test equipment, if the device under test 
is charged. Alternatively connect the tester to a working power supply or dis-
charge the device under test with the ground and discharge rod.

 

Содержание S-110 DC

Страница 1: ......

Страница 2: ......

Страница 3: ...USER MANUAL HIGH VOLTAGE INSULATION TESTERS S 50 DC S 80 DC S 110 DC S 120 DC SONEL S A Wokulskiego 11 58 100 widnica Version 1 02 11 09 2020...

Страница 4: ...S 120 DC USER MANUAL 2 S 50 80 110 120 DC is a modern easy in use and safe testing device Please acquaint yourself with this manual in order to avoid measuring errors and prevent possible problems in...

Страница 5: ...g the discharge rod 8 3 4 Attaching the corona ring 9 3 5 Connecting the test equipment 10 4 Operating the device 13 4 1 Controls and displays 13 4 2 Operating procedure 14 4 2 1 Setting the operating...

Страница 6: ...th Only operate the system if it is in technically perfect condition No non original parts may be used for the system and its peripherals as the necessary safety will not be guaranteed No mode of work...

Страница 7: ...e equation I U R where U measuring voltage I measured current A special protective ground circuit ensures a high level of safety 2 1 Design of the set The test set consists of the operating unit and t...

Страница 8: ...80 110 120 DC is used for electrical testing of power cables and plant with direct voltage Areas of application include testing at re commissioning of cable systems after maintenance and repair testin...

Страница 9: ...ting unit 3 2 Operating requirements The carrying handle of the operating unit is designed for one hand operating After putting down the device carrying handle can be locked in 30 steps by using a pul...

Страница 10: ...ve instructions it must be placed in close proximity to the test object so that the ground and high voltage cables can be kept as short as possible When the equipment is set up in the field it must be...

Страница 11: ...spacer is only plugged onto the HV unit without screwing Connect the HV lug only hand tight with the knurled screw Consider following differences to the standard design The smoothing of electricals fi...

Страница 12: ...tablish a conductive connection between the station auxiliary ground and the ground terminal of the high voltage unit Screw the high voltage attachment onto the high voltage unit and connect the high...

Страница 13: ...gle core system or a three core cable with separate shield on each core All three cores can be tested at the same time Fig 3 4 Connecting the test equipment for insulation testing of a three core cabl...

Страница 14: ...ANUAL 12 Fig 3 5 Connecting the test equipment for sheath testing of a 3x single core system The sheaths of all three cores can be tested at the same time The cable shields at the other end of the cab...

Страница 15: ...f it is in the leftmost position while enabling the button Unit on lamp white Lights up when the unit is switched on with the mains switch and mains voltage is connected Ground terminal lamp yellow Li...

Страница 16: ...ble 1 ground cable for operating unit to station ground Pilot lamp signals proper connection of the station and protective grounds independently of the high voltage being connected The equipment will...

Страница 17: ...s exceeded by approx 5 kV the overvoltage trip will re spond If this happens the high voltage will have to be switched back on set the high voltage adjuster to zero first Fig 4 2 Left meter display Ri...

Страница 18: ...t must only be operat ed in areas which have been secured and or cordoned off and marked in accordance with EN 50191 and EN 50110 1 standards In an emergency the high voltage must be switched off by a...

Страница 19: ...re external soiling direct sunlight and ambient temperatures in excess of 55 C must be avoided NOTE The high voltage unit must only be transported in an upright position 6 Technical specifications a p...

Страница 20: ...18 5 kg S 110 DC 20 0 kg S 120 DC 20 5 kg o degree of protection as per EN 60529 operating unit IP40 high voltage unit IP65 high voltage unit at terminals IP00 p max voltage discharge energy for disc...

Страница 21: ...e 3 blue lead for high voltage unit length 2 m cable 4 ground cable for discharge rod discharge rod case for discharge rod user manual service kit packing 7 2 Optional accessories Additionally the fol...

Страница 22: ...ermal imagers o luxmeters The Calibration Certificate is a document that presents a relation between the calibration standard of known accuracy and meter indications with associated measurement uncert...

Страница 23: ......

Страница 24: ......

Отзывы: