OPERATING MANUAL PAT-810/815/820 version 1.06
81
adjustable measuring time
in the range of: 1 s...60 s, with resolution of 1 s and option
Continuous test
(enabled by default), in AUTOTEST only adjustable: 1...60 s, with resolution
of 1 s
* only for current measurement with clamp
(8 % m.v. + 5 digits)
PF Power Factor
Display range
Resolution
Basic uncertainty
0,00…1,00
0,01
± (10% m.v. + 5 digits)
adjustable measuring time
in the range of: 1 s...60 s, with resolution of 1 s and option
Continuous test
(enabled by default), in AUTOTEST only adjustable: 1...60 s, with resolution
of 1 s
Current Consumption measurement during power measurement
Display range
Resolution
Basic uncertainty
0,00 A...15,99 A
0,01 A
(2 % m.v. + 3 digits)
adjustable measuring time
in the range of: 1 s...60 s, with resolution of 1 s and option
Continuous test
(enabled by default), in AUTOTEST only adjustable: 1...60 s, with resolution
of 1 s
Current Consumption measurement with clamp during power measurement
Display range
Resolution
Basic uncertainty
100 mA…999 mA
1 mA
(5 % m.v. + 5 digits)
1,00 A…9,99 A
0,01 A
10,0 A…2
4,9 A
0,1 A
basic uncertainty above does not include uncertainty of measuremenr clamp
adjustable measuring time
in the range of: 1 s...60 s, with resolution of 1 s and option
Continuous test
(enabled by default), in AUTOTEST only adjustable: 1...60 s, with resolution
of 1 s
Voltage measurement in test socket
Display range
Resolution
Basic uncertainty
195,0 V…265,0 V
0,1 V
(2 % m.v. + 2 digits)
Measurement of RCD parameters
RCD tripping time measurement for sine t
A
differential current
Measurement range in acc. with IEC 61557: 0 ms ... up to the upper limit of displayed value
RCD type
Rated Current
Multiplication
Factor
Measurement
range
Resolution
Basic uncertainty
General
0,5 I
n
0..300 ms
1 ms
± 2
% m.v.
±
2 digits
1)
1 I
n
2 I
n
0..150 ms
5 I
n
0..40 ms
1)
for I
n
= 10 mA and 0,5 I
n
uncertainty is
± 2
% m. v.
±
3 digits